52
Error Codes
Display
Cause
Solution
E 22
Weight is too light or there is no
Increase the weight on the balance/
sample on the balance/scale
scale
E 30
„Print“ key pressed while balance/
Activate SBI mode (e.g., using the
scale is in BPI mode
„Close“ function in the service software)
E 50
Overload or underload in temperature Defect in TK sensor or main PCB
compensation converter
(replace)
E 53
Temperature compensation converter
Defect in TK sensor or main PCB
not functioning. No measured value
(replace)
received in processor from TK switch
E 54
A/D converter output is below the
Device is underloaded; no weighing
minimum level
pan/ load plate on balance/scale;
mechanical defect; defective main PCB
E 55
A/D converter output is above the
Device is overloaded; mechanical
maximum level
defect; defective main PCB
Summary of Contents for Talent Series