27
Anz-019F.eps
The equipment is now connected as follows:
– The balance/scale has a 25-pin RS-232 interface for the service software
– There is a power supply to the balance/scale (beside the battery, if one is
installed) so that the BPI mode can be activated.
– Replace the weighing pan/load plate and turn on the balance/scale.
– When 0.0 is displayed, use a suitable tool to press and hold the button on
the BPI adapter (accessible through an opening in the housing.
– The balance/scale now performs its self-test (display segment check)
repeatedly.
– Wait until the segment check has lit up at least 3 times; then press the
button again.
– The balance/scale electronics are now in the BPI mode and you can use the
balance/scale with the service software.
Caution!
After working in the BPI mode, make sure to set the write-
protect again (using the „Close“ function in the service
software program) so that the balance/scale returns to the SBI
mode (SBI = serial balance interface; the standard data
transfer protocol). Otherwise one will not be able to operate
peripheral devices with the balance/scale; when
r
is pressed,
the error code „E 30“ is displayed.