26
Aut12420.jpg
Activating the BPI Mode
Note:
The BPI mode must be activated in the balance/scale before you can work with
the service software. You need to use this software, for example, to adjust the
linearity and/or span after adjusting the zero-point offset value, or after
replacing a PCB.
The BPI adapter (part no. 6740-71) must be connected to the balance/scale
before the BPI mode can be activated. The procedure for connecting the
adapter is as follows:
– Turn off the balance/scale and disconnect it from the power supply
– Remove the weighing pan/load plate
– Remove the protective cap (A) from the bottom of the balance/scale.
– The main PCB is now accessible through an opening in the balance/scale
housing.
– Connect the white plug on the cable of BPI adapter to the corresponding
position on the main PCB.
– Connect the plug on the power cable of the BPI adapter to the power jack
on the balance/scale.
– Plug the balance/scale AC adapter into the power jack on the BPI adapter.
BPI-Button
Anschluss
BPI02.jpg
BPI-Button
Main PCB