Opening and Closing the Balance/Scale
Opening the Balance/Scale
– Disconnect the balance/scale from power (remove the battery, if one is
installed).
– Remove the weighing pan/load plate, as well as any shock absorbers and/or
pan support (B), and place to one side.
– Remove the protective caps (A) and remove the Phillips screws.
– Remove the upper part of the housing by carefully pulling it towards the
front. Make sure that you do not damage the ribbon cable that connects the
display PCB to the main PCB.
– The display PCB, main PCB and strain-gauge load receptor are now accessible.
Closing the Balance/Scale
– Connect any internal lines that were unplugged.
– Slide the upper part of the housing into position and tighten the Phillips
screws.
– Place the weighing pan/load plate, as well as any shock absorber and/or pan
support (B), on the balance/scale.
– Connect the equipment to power.
– Check all balance/scale functions.
– Replace the protective caps over the Phillips screws when all other steps have
been completed.
Important:
After repairing and testing the balance/scale, make sure to
attach your company seal (C).