
29
Checking and Adjustment Sequence
Checking the Overload Stop
Note:
Overload stops must be checked only following replacement of the strain-
gauge system.
The overload stop must be reset after the strain-gauge load receptor has been
damaged or replaced. Weight readouts greater than the balance/scale’s
maximum capacity must be read off. The service software is required for this
procedure.
– Activate the BPI mode (see page 26)
– Service software: Select the „Diagnostics“ program; then select the „High
Resolution“ subprogram
– When adjusting the overload stop, read off the weight value shown on the
Psion server or PC display (see page 30)
– After adjusting the overload stop, activate the „Close“ function in the service
software to reset the balance/scale to the SBI mode.
Summary of Contents for Talent Series
Page 55: ...55 ...