102
Specifications
Secura, Quintix User Manual
Quintix
®
Model
125D-1x
1
)
65-1x
1
)
35-1x
1
)
Accuracy class
k
k
k
Type
2
)
SQP-F
SQP-F
SQP-F
Max
g
60 / 120
60
30
Scale interval d
g
0.00001 / 0.0001
0.00001
0.00001
Verification scale interval e
g
0.001
0.001
0.001
Min
g
0.001
0.001
0.001
Min (only for Models ...-1OIN)
g
0.1
0.1
0.1
Tare equalization range (subtractive)
< 100% from max. weighing capacity
Typical stabilization time
s
6 / 2
6
6
isoCAL:
– Temperature change
– Time interval
K
h
1.5
4
1.5
4
1.5
4
Display result
(depending on the set filter level)
s
0.2 / 0.4
0.2 / 0.4
0.2 / 0.4
Weighing pan size
mm
d
80 (optional
d
90)
d
80 (optional
d
90)
d
80 (optional
d
90)
Weighing chamber height*
mm
218
218
218
Net weight, approx.
kg
7.8
7.8
7.8
1
) Possible terms for country-specific models:
x = CEU: Verified balances with EC Type Approval Certificate D12-09-014 without country-specific additions
x = CFR: Verified balances with EC Type Approval Certificate D12-09-014 for France only
x = CIT: Verified balances with EC Type Approval Certificate D12-09-014 for Italy only
x = CCH: Verified balances with EC Type Approval Certificate D12-09-014 for Switzerland only
x = CN: CMC Type Approval Certificate for China
x = OJP: Balance with Type Approval Certificate for Japan
x = OBR: Balance with Type Approval Certificate for Brazil
x = ORU: Balance with Type Approval Certificate for Russia
x = OIN: Balance with Type Approval Certificate for India
2
) all Models ...CN: Type “SQP”
*
Upper edge of the weighing pan to the lower edge of the upper draft shield panel
Visit www.scaleman.com your Authorized Sartorius Distributor
Visit www.balances.com for the full line of Sartorius Laboratory Balances