Calibration and Adjustment
Secura, Quintix User Manual
8.2 Calibration/Adjustment Using an External Cali-
bration Weight
An external calibration weight is required for this function.
Please note the tolerance of the calibration weight being used.
In verified balances with accuracy class
k
, external adjustment can only be accessed
when the access switch is open.
t
Make sure that the weighing pan is empty.
t
Select
A
to zero the balance.
t
Select the menu key at the bottom left of the display in the current application.
Components
Statistics
Mixing
Weighing
y
The menu appears.
t
Select
the
CAL
button to access the calibration function.
Calibrate balance
CAL-Extern
CAL-Intern
y
The
Calibrate balance
window appears.
t
Select
CAL-Extern
.
y
A message appears on the display, prompting the user to place the calibration
weight on the weighing pan.
t
Place the calibration weight on the weighing pan.
y
Calibration starts automatically.
Date and Time:
14-Jun-2014 14:13
Calibration function:
External calibration
Start calibration:
Start: manually
Calibration weight:
Set 100.00000 g
Calibration deviation:
Dev 0.00002 g
Adjustment function:
External adjustment
Adjustment deviation:
Dev 0.00000 g
Report
y
A report appears on the display.
The report indicates the deviations identified during calibration. The report also
gives the result of the adjustment process.
t
Close the report window: Select
<
.
y
The balance is now calibrated and adjusted.
Information on potential errors can be found in Chapter “Status Messages” on
page 90.