51
Weighing
Secura, Quintix User Manual
7.6 Totalizing
Components
Purpose: Totalize weight values. You can save up to 99 components.
This allows you to save components that must be weighed in various containers.
Each container can be tared before each component is weighed.
Example: You have a formula, the components of which should not be mixed together
immediately. Use this application to weigh individual components in different
containers, print out the results, and check the total weight at any time.
t
Select the menu key in any application.
t
Select
the
Components
symbol in the menu.
1
0.00000
Max 120 g
d=0.00001 g
Component
y
The
Components
application appears.
Component
1
t
To change the settings, select the gray button.
Components
Print component
On
Off
y
The
Components
settings window appears.
t
Choose whether the individual weights of the components are printed out.
t
Touch
l
to confirm.
t
Select
A
to zero the balance if necessary.
t
Place the container for the first component on the weighing pan.
t
Select
D
to tare the balance.