75
Calibration and Adjustment
Secura, Quintix User Manual
8.3
Automatic Calibration/Adjustment with isoCAL
Fully automatic calibration/adjustment ensures that the balance is automatically
calibrated and adjusted when predefined time intervals or temperature values are
exceeded.
This function varies in different models:
–
isoCAL is automatically carried out based on time or temperature, and each time
the balance is leveled. This function can be switched off via the system settings
(see page 30).
In
Secura
®
models, every adjustment process, including the identified deviations,
is documented and saved on the balance. The saved reports can be viewed via the
system settings and printed out on a PC or Sartorius lab printer (see page 30).
0.00000
Max 120 g
d=0.00001 g
isoCAL
y
If the ambient temperature has changed since the last calibration/adjustment or
the preset time interval has been exceeded, the red
isoCAL
button appears on
the display.
y
If the
isoCAL
function is set to
Automatic
, the calibration/adjustment process
starts automatically.
!
If the balance has not been adjusted, the prompt will be shown again 5 minutes
after the “isoCAL” status message is acknowledged.
y
If
the
isoCAL
function is set to
Info, manual start
, the function must be started
manually.
isoCAL
t
To start the calibration/adjustment function manually, select the
isoCAL
button.
y
Calibration starts automatically.
Date and Time:
14-Jun-2014 15:18
Calibration function:
Internal calibration
Start calibration:
Start: isoCAL (Niv)
Calibration deviation:
Dev -0.00001 g
Adjustment function:
Internal adjustment
Adjustment deviation:
Dev 0.00000 g
Report
y
A report appears on the display.
The report indicates the deviations identified during isoCAL calibration.
The report also gives the result of the adjustment process.
t
Close the report window: Select
<
.
y
The balance is now calibrated and adjusted.