29
defective ballasts)
Electrical wires are disconnected
Plug them properly
One or more relay is disconnected
Connect the relay(s) properly
One relay is defective
Contact SANUVOX (replace the
relay)
5.
The cooling fan doesn’t
work.
Electrical wires are disconnected
Plug them properly
The relay is loosely connected to
the board
Push on the relay to have a tight
connection
Relay’s pins are twisted
Straighten up the pins and firmly
connect the relay.
The relay is defective
Contact SANUVOX (replace the
relay
6.
Only the primary tower
works, not the secondary
tower.
The unit is in « Primary unit only »
mode
In the ADMIN mode on the
tablet, unselect « Primary unit
only »
7.
The lamps don’t shut off
when someone walks in.
One or more motion detectors are
sunk in: the visual filed may be
reduced
Try to fix the motion sensors in a
protruding position
One or more motion detectors are
defective
Contact SANUVOX (replace the
defective sensors)
8.
Cannot access the web
interface.
Disconnected Ethernet cable
between PLC and router
Connect properly the cable
Electrical wires are disconnected
Connect properly the wires
The PLC is defective
Contact SANUVOX (replace or
reprogram the PLC)