Caution
The number flashes for about 10 seconds.
Press ▲ or ▼ key during the period
of 10 seconds flashing. Otherwise the buzzar
beeps, beeps to return automatically to the
beginning. Try again to reselect the number
from the beginning. In case ■ or ▼ is
indicated on the right side (* marked) of
Cal.No, calibrated values are stored in
the calibration.
When no more use for the value, process
zeroing /calibration and delete the
stored data that is replaced with new data.
3) How to select 「Calibration No: 」
It is capable of getting started on measurements immediately after by switching Power to ON as described
on page10, 「it becomes possible to take measurements and adjusting」.
However, it may make errors depending on material formation and shapes to be measured. To minimize
measurement errors and obtain as accurate results as possible please be sure of carrying out 2 points
adjustments of 「Zeroing」 and 「Calibration standard」 before measuring process.
And this set of being adjusted is called
“
Calibration
”
SWT-7200Ⅱ is capable of storing 10 pares of calibration data.
The storing place is indicated by a calibration No.
「NFe」 is indicated when a SWT exclusive probe in
NFe series is connected.
Press CAL No. key.
The buzzer emits a beeping sound.
A number on the display flashes.
Press ▲ key or ▼ key to select your
desired number.
-14-
(Fe)
-
μm
Cal.No. 1
(Fe)
-
μm
Cal.No. 1