Function switching
(1)Switching to Noninterrupt Measurement Mode
Switch to「non-interrupt measurement mode」when
it is necessary to slide a probe along the measuring
surface of a substrate as illustrated on the right figure
for continuous measurements of films.
A measuring value is indicated and stored
each time a probe is pressed in a normal state.
Hold ZERO key and press ▲ key.
The buzzer beeps, beeps.
This unit has turned into 「noninterrupt measurement mode」. Measurements can be made about in
0.5 seconds intervals and the data is indicated with a beeping sound.
↓ ↑
Traces of probe
↓ ↑
→ → → → → →
- 751
- 751
837 125 124 126 123 124 751
Measuring values on display(indicated with blinking successively each 0.5 seconds interval)
※
The non-interrupt function is stored when switching Power to OFF, and can be maintained until
re-activating to switch to ON.
-30-
Substrate
Substrate
(Fe)
125
μm
Cal. No. 1 ■