Table 3
Guidance and manufacturer’s declaration – electromagnetic immunity
The SEM 42 is intended for use in the electromagnetic environment specified below. The customer or the user of the SEM 42 should assure that it is used in such an envi-
ronment.
IMMUnITY test
IEC 60601 test level
Compliance level
Electromagnetic environment – guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic tile. If floors are covered
with synthetic material, the relative humidity should be at least 30 %.
Electrical fast transient/burst
IEC 61000-4-4
± 2 kV for power supply lines
± 1 kV for input/output lines ± 1 kV for input/output lines
Mains power quality should be that of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
± 1 kV line to line
± 2 kV line to earth
Not applicable
Mains power quality should be that of a typical commercial or hospital
environment.
Voltage dips, short interruptions and
voltage variations on power supply
input lines
IEC 61000-4-11
<5 % U
T
(>95 % dip in U
T
)
for 0.5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5 s
Not applicable
Mains power quality should be that of a typical commercial or hospital
environment. If the user of the SEM 42 requires continued operation
during power mains interruptions, it is recommended that the SEM 42
be powered from an uninterruptible power supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at levels characteristic of a
typical location in a typical commercial or hospital environment.
nOTE:
U
T
is the a.c. mains voltage prior to application of the test level.
Summary of Contents for SEM 42
Page 18: ...18 2 EMS TENS 39 C EMS TENS 1 m...
Page 19: ...19 9 mA cm 2 mA cm piercing piercing 5 10 3...
Page 20: ...20 3 1 EMS TENS 3 2 Hz Hertz 15 Hz 35 Hz 45 70 Hz 3 3 3 4 3 5 EMS TENS EMS TENS 6...
Page 28: ...28 9 Pb Cd Hg 2002 96 Waste Electrical and Electronic Equipment 10 1 2 3 1 2 1 2 3 4 5 6 1...