![Samwontech TEMI770 Manual Download Page 67](http://html1.mh-extra.com/html/samwontech/temi770/temi770_manual_3793449067.webp)
III : COMMUNICATION
IM TEMI 770E_C 2nd Edition: Sep. 1. 2003. 8/21
3.7 WRD Command
This command for writing D-Register random.
■
Transmission Format
Byte 1 2
3
1 2 1
4
1
4
1
…
S
T
Command
element
X
Addr WRD ,
Param
e
ter
numb
e
r
,
D-Reg. No1
,
dddd-1
,
…
1 4 1 4 2
1
1
C
L
, D-Reg.
No(n) ,
dddd-(n)
SUM
R
F
■
Response
Byte
1
2
3 1
2 2
1
1
S C
L
T
Command
element
X
Addr WRD ,
OK
SUM
R
F
- Parameter number : 1 ~ 32
- dddd : Indicates a character string in hexadecimal format.
ex) When the writing Temp SP(D0102) and Humi SP(D0106) as the FIX Running.
- Setting Temp SP : 50.0
℃→
Delete decimal point(500)
→
Hexadecimal(0x01F4)
- Setting Temp SLOP : 0.5
℃→
Delete decimal point(5)
→
Hexadecimal(0x0005)
- Transmission : [stx]01WRD,02,0102,01F4,0106,0005[cr][lf]
- Transmission(Include Check Sum) : [stx]01WRD,02,0102,01F4,0106,0005B6[cr][lf]