IM TEMI 770E_C 2nd Edition: Sep. 1. 2003. 34/40
5-6-6. ALARM
Select
ALARM
at the SETUP page(Figure 5-6-1) for switching the ALARM page.
Figure 5-6-13 ALARM
①
ITEM
: Alarm mode.(Temp/Humi)
②
TYPE
: Alarm type.
③
POINT
: Alarm value.(-100~100%)
④
HYSTERESIS
: Set HYSTERESIS on Alarm.
Alarm is still working even if on STOP mode.
(Table 5-14) ALARM PARAMETER
Name
Function
Valid Range
Unit
Default Value
ITEM ITEM TEMP,
HUMI
ABS
TEMP
TYPE
ALARM TYPE
1 ~ 20(See Figure 5-15)
ABS
1
POINT
ALARM POINT
EUS(-100 ~ 100%)
EU
EU(0%)
HYS ALARM
HYSTERESIS
EUS(0 ~ 100%)
EUS
EUS(0.5%)
Switching