STEP:
1 step mode (once mode)
LOOP:
loop mode
Return command: none
Example:
If sending command
[:]LIST:MODE CONT
, it means the working mode of list files is set as continuous mode.
2. Related command
[:]LIST:MODE?
, it means to query the working mode of the current list.
Return parameters: <mode>
Command [:]LIST:VOLTage <count>,<value>
1. This command is used to set the setup voltage value of the appointed step in the list.
Parameter: count is appointed step number; value is the setup voltage.
Unit: V
Return command: none
Example:
If sending command
[:]LI
ST:VOLT 1,5
, it means to set the setup voltage value of the first step of the list file
as 5.000V.
2. Related command
[:]LIST:VOLTage? <count
>
, it means to query the setup voltage value of the appointed
step in the list.
Return parameters:<value>
Example:
If sending command
[:]LIST:VOLT? 1
, it means to query the setup voltage value of the first step in the current
list.
Command [:]LIST:CURRent <count>,<value>
1.This command is used to setup current value of the appointed step in the list.
Parameters:
count
is appointed step number;
value
is the setup current.
Unit: A
Return command: none
Example:
If sending command
[:]LIST:CURR 1,2
, it means to set the setup current value of the first step in the list as
2.000A.
2. Related command
[:]LIST:CURRent? <count>
, it means to query the setup current value of the appointed
step in the list.
Return parameter: <value>
Example:
If sending command
[:]LIST:CURR? 1
, it means to query the setup current value of the first step in the current
list.