4.
CARE OF THE INSTRUMENT
Page 26
List of Disinfection/Sterilization Methods
CAUTION: When sterilizing the guard bowl unit by
high pressure steam, wear protective clothing like
gloves and take care not to get burned.
CAUTION: All of the components of the guard bowl
are autoclaveable. Each component must be
sterilized separately.
CAUTION: If the bowl lid, rotor and guard bowl are
placed in the same container for disinfection, they
may be deteriorated faster (dissimilar metal
corrosion). It is recommended to disinfect them
separately.
4.1.4 Cleaning 6mL/12mL specimen
chambers, 6mL/12mL rubber
gaskets, 6mL/12mL chamber cap
and chamber holder
(Figure 4M)
These assemblies consist of the following four parts:
1 specimen chamber holder
2 rubber gasket
3 specimen chamber
4 specimen chamber cap
Figure 4M
All these components should be carefully decontaminated
and cleaned prior to use.
1.
All components should be soaked in an approved
disinfectant. A 0.5% chlorhexidine hydrochloride
solution or a 0.5% or less sodium hypochlorite solution
(common bleach) are the recommended disinfectants.
2.
After soaking the components in an approved
disinfectant, they should be cleaned in hot water and
Component Disinfection/Sterilization
Bowl Lid
Can be sterilized in an autoclave (121°C, 20
minutes).
Can be disinfected by soaking in ethyl alcohol
(80%), isopropyl alcohol (70%) or sodium
hypochlorite solution (0.5% or less in
concentration, for less than 20 minutes).
Rotor
Can be sterilized in an autoclave (121°C, 20
minutes).
Can be disinfected by soaking in ethyl alcohol
(80%), isopropyl alcohol (70%) or sodium
hypochlorite solution (0.5% or less in
concentration, for less than 20 minutes).
Guard Bowl
Can be sterilized in an autoclave (121°C, 20
minutes).
Can be disinfected by soaking in ethyl alcohol
(80%), isopropyl alcohol (70%) or sodium
hypochlorite solution (0.5% or less in
concentration, for less than 20 minutes).
CAUTION:
If the guard bowl is autoclaved as it is
moistened with sodium hypochlorite
solution, it will be wear down faster
.
Bowl Lid
Gasket
Can be sterilized in an autoclave (121°C, 20
minutes).
Can be disinfected by soaking in ethyl alcohol
(80%), isopropyl alcohol (70%) or sodium
hypochlorite solution (0.5% or less in
concentration, for less than 20 minutes).
V-shape
Gasket