CARE OF THE INSTRUMENT
Page 23
4.1 General Information
The Cyto-Tek 2500 Cytocentrifuge is a precision instrument
designed for trouble-free operation and requires minimal
routine maintenance. Handle the instrument with care to
avoid damage to the sensitive electronics. The
cytocentrifuge may be kept in good operating condition by
carefully following these instructions. Before cleaning or
performing maintenance on the instrument, place the ON-
OFF Switch in the OFF position and unplug the power cord
from its electrical outlet.
WARNING: WHEN PERFORMING MAINTENANCE OR
CLEANING THE INSTRUMENT, STANDARD
LABORATORY SAFETY PROCEDURES AND PERSONAL
PROTECTIVE EQUIPMENT SHOULD BE USED.
4.1.1 Cleaning the Control Panel
Wipe the internal surfaces (including the inside lid area and
the well) with a clean cloth moistened with neutral
detergent. Dry any excessively moist areas.
CAUTION: Do not use volatile liquids such as
solvents.
4.1.2 Cleaning the Instrument
Wipe the external surfaces (Lid Gasket area and Guard
Bowl Storage area) with a clean cloth moistened with 70-
80% ethyl alcohol or isopropyl alcohol for disinfection. Dry
any excessively moist areas.
CAUTION: Do not spray a disinfectant directly on or
over the instrument.
4.1.3 Disassembling and Cleaning
the Guard Bowl Unit
CAUTION: To lessen the infection risk, place the
guard bowl unit inside a safety cabinet before
removing and the lid.
To clean, disinfect, or sterilize the guard bowl unit, it is
necessary to place and disassemble it inside a safety
cabinet. Remove the lid gasket from the lid and the rotor
unit and the V-shape gasket from the guard bowl before
cleaning.
Disassembly of the bowl lid
1.
Remove the bowl lid (Figure 4A).
Figure 4A
2.
Remove the lid gasket (Figure 4B).
Figure 4B