24–40 GHz SPECTRUM COMPACT v.2 (FW V5.3.3)
OPERATION
21
2.10
SAVE
Tapping the SAVE button instantly saves the spectrum curve with an automatically
generated file name: the first letter indicates the trace mode (“C” corresponds to
NORMAL trace, “M” to MAXHOLD trace, “W” to CUMULATIVE trace, “A” to AVERAGE
2/4/8/16 traces and “N” to MIN/MAXHOLD trace) and the remaining digits are the file
order number.
Holding the SAVE button allows saving the current spectrum curve with a custom
file name. The name can be up to 8 characters long.
When assigning a custom file name, it is checked whether the name does not
match with an already existing one. In that case it is not possible to confirm.
SAVE button is not available if at least one full sweep has not been done with
the current sweep settings.
After saving, the curve is displayed with the file name in the upper left corner of the
graph. The saved spectrum curves can be reviewed at any time at TOOLS & SETTINGS
FILES.
Saving a spectrum curve stops continuous sweep and puts the saved file on screen.
To resume continuous sweep (thus closing the file), tap SWEEP CONT. button.
The SAVE button
File with automatically assigned name