Chapter 4: Patient preparation
34
Connection of the Chest Electrodes
Press the suction bulb and place the chest electrode on proper site (Figure 4-3) and then release.
The location of the chest electrodes is as follows:
C1 (V1): Fourth intercostal space at the right margin of the sternum
C2 (V2): Fourth intercostal space at the left margin of the sternum
C3 (V3): Midway between V2 and V4
C4 (V4): Fifth intercostal space at the left midclavicular line
C5 (V5): Left anterior axillary line at the horizontal level of V4
C6 (V6): Left mid-axillary line at the horizontal level of V4
Detection of electrode disconnection
DENA 650 continuously monitors the connection status of the electrodes, and in the event of a
disconnection, displays the relevant messages at the location of the signals on the screen (Figure 1-2).
If any of the R, L, or F electrodes are disconnected, the messages Check R, Check L, and Check F
are displayed.
If any of the chest electrodes are disconnected, the message Check Cx is displayed (x from 1 to
6).
If the N electrode is disconnected, one or more disconnection messages may be displayed.
NOTE
Refer to Troubleshooting and Error messages
chapter to view device messages.
Figure 4-3
Connection of the Chest Electrodes
Summary of Contents for Dena 650
Page 1: ...User Manual Dena Electrocardiograph DENA 650 Linux AADAT Co D00912 7...
Page 3: ......
Page 11: ......
Page 12: ...Chapter 1 Introduction 1 1 Introduction...
Page 25: ...Chapter 1 Introduction 14...
Page 26: ...Chapter 2 System settings 15 2 System settings...
Page 38: ...Chapter 3 Patient information 27 3 Patient information...
Page 41: ...Chapter 3 Patient information 30...
Page 42: ...Chapter 4 Patient preparation 31 4 Patient preparation...
Page 47: ...Chapter 4 Patient preparation 36 Lead Placement Diagram Figure 4 4 Lead placement digram...
Page 48: ...Chapter 5 Recording Operation 37 5 Recording Operation...
Page 52: ...Chapter 6 ECG analysis and measurement 41 6 ECG analysis and measurement...
Page 58: ...Chapter 7 Data management 47 7 Data management...
Page 64: ...Chapter 8 Care and Cleaning 53 8 Care and Cleaning...
Page 71: ...Chapter 8 Care and Cleaning 60...
Page 72: ...Chapter 9 Troubleshooting and Error messages 61 9 Troubleshooting and Error messages...
Page 77: ...Chapter 9 Troubleshooting and Error messages 66...
Page 78: ...Chapter 10 Technical specification 67 10 Technical specification...
Page 81: ...Chapter 10 Technical specification 70...
Page 82: ...Appendix 1 Accessories 71 Appendix 1 Accessories...
Page 84: ...Appendix 2 System parameters 73 Appendix 2 system parameters...
Page 87: ...Appendix 2 System parameters 76...
Page 88: ...Appendix 3 Electromagnetic compliance EMC 77 Appendix 3 Electro magnetic compliance...
Page 93: ...Appendix 3 Electromagnetic compliance EMC 82...
Page 94: ...Appendix 4 The GLASGOW program 83 Appendix 4 The GLASGOW program...
Page 101: ......