Chapter 2: System settings
22
User Setting Menu
Select
User
from the Main Menu to access the below menu:
The following items can be set through this menu:
Save: Enable this item to save patient data and signals in Auto and Rhythm modes.
Pace Detect: Available options are Off and On. DENA 650 detects and rejects pacemaker-generated
signals from ECG signal, so that they will be ignored in determining heart rate. If you select On for
patients with pacemaker, detected pacemaker signals will be marked on the ECG waveform as a
white vertical line.
Measurement: If you select Global or Details, a measurement table will be printed at the end of
recording. For more information, refer to ECG analysis and measurement chapter.
WARNING
For patients with pacemaker, the PACE DETECT function must be switched "ON", otherwise,
the pacing impulse may be counted as normal QRS complex.
In patients with pacemaker, if the PACE DETECT function is "OFF", turn off the low pass
(25Hz, 35Hz) and EMG filters to check pacemaker function.
Figure 2-5
User setting
Summary of Contents for Dena 650
Page 1: ...User Manual Dena Electrocardiograph DENA 650 Linux AADAT Co D00912 7...
Page 3: ......
Page 11: ......
Page 12: ...Chapter 1 Introduction 1 1 Introduction...
Page 25: ...Chapter 1 Introduction 14...
Page 26: ...Chapter 2 System settings 15 2 System settings...
Page 38: ...Chapter 3 Patient information 27 3 Patient information...
Page 41: ...Chapter 3 Patient information 30...
Page 42: ...Chapter 4 Patient preparation 31 4 Patient preparation...
Page 47: ...Chapter 4 Patient preparation 36 Lead Placement Diagram Figure 4 4 Lead placement digram...
Page 48: ...Chapter 5 Recording Operation 37 5 Recording Operation...
Page 52: ...Chapter 6 ECG analysis and measurement 41 6 ECG analysis and measurement...
Page 58: ...Chapter 7 Data management 47 7 Data management...
Page 64: ...Chapter 8 Care and Cleaning 53 8 Care and Cleaning...
Page 71: ...Chapter 8 Care and Cleaning 60...
Page 72: ...Chapter 9 Troubleshooting and Error messages 61 9 Troubleshooting and Error messages...
Page 77: ...Chapter 9 Troubleshooting and Error messages 66...
Page 78: ...Chapter 10 Technical specification 67 10 Technical specification...
Page 81: ...Chapter 10 Technical specification 70...
Page 82: ...Appendix 1 Accessories 71 Appendix 1 Accessories...
Page 84: ...Appendix 2 System parameters 73 Appendix 2 system parameters...
Page 87: ...Appendix 2 System parameters 76...
Page 88: ...Appendix 3 Electromagnetic compliance EMC 77 Appendix 3 Electro magnetic compliance...
Page 93: ...Appendix 3 Electromagnetic compliance EMC 82...
Page 94: ...Appendix 4 The GLASGOW program 83 Appendix 4 The GLASGOW program...
Page 101: ......