
Chapter 4: Patient preparation
33
Connection of theLimb Electrodes
4 electrodes of 10 ECG electrodes are attached to the limbs.
The location of the limb electrodes for the 12-lead ECG is as follows:
Left hand (L)
Right hand (R)
Left foot (F)
Righ foot (N)
WARNING
Use only one type of electrode on the same patient to avoid variations in electrical impedance.
It is recommended to use silver/silver chloride electrode. When dissimilar metals are used for
different electrodes, the electrodes may cause large offset potentials due to polarization, which
may be severe enough to prevent obtaining an ECG trace. Using dissimilar metals may also
increase recovery time after defibrillation.
Use intact and clean electrodes only. Electrodes with damaged surface may cause ECG
waveform inaccuracy.
Use only the manufacturer recommended ECG cable with internal resistance. Other ECG cables
and leads may cause burns, improper performance and/or provide inadequate protection during
defibrillation.
When you connect the cables and electrodes, make sure that no metal part is in contact with the
safety ground.
Verify that all ECG electrodes are correctly attached to the patient.
Interference from a non-grounded instrument near the patient and/or ESU (Electrosurgical Unit)
interference can cause inaccuracy of ECG waveform.
When the device is used with electrocautery unit, please note the position of leads. In order to
reduce the hazard of burns, the leads should be located away from the electrocautery pen and
return electrode.
..دنا هدش
Figure 4-2
Connection of the Limb Electrodes
Summary of Contents for Dena 650
Page 1: ...User Manual Dena Electrocardiograph DENA 650 Linux AADAT Co D00912 7...
Page 3: ......
Page 11: ......
Page 12: ...Chapter 1 Introduction 1 1 Introduction...
Page 25: ...Chapter 1 Introduction 14...
Page 26: ...Chapter 2 System settings 15 2 System settings...
Page 38: ...Chapter 3 Patient information 27 3 Patient information...
Page 41: ...Chapter 3 Patient information 30...
Page 42: ...Chapter 4 Patient preparation 31 4 Patient preparation...
Page 47: ...Chapter 4 Patient preparation 36 Lead Placement Diagram Figure 4 4 Lead placement digram...
Page 48: ...Chapter 5 Recording Operation 37 5 Recording Operation...
Page 52: ...Chapter 6 ECG analysis and measurement 41 6 ECG analysis and measurement...
Page 58: ...Chapter 7 Data management 47 7 Data management...
Page 64: ...Chapter 8 Care and Cleaning 53 8 Care and Cleaning...
Page 71: ...Chapter 8 Care and Cleaning 60...
Page 72: ...Chapter 9 Troubleshooting and Error messages 61 9 Troubleshooting and Error messages...
Page 77: ...Chapter 9 Troubleshooting and Error messages 66...
Page 78: ...Chapter 10 Technical specification 67 10 Technical specification...
Page 81: ...Chapter 10 Technical specification 70...
Page 82: ...Appendix 1 Accessories 71 Appendix 1 Accessories...
Page 84: ...Appendix 2 System parameters 73 Appendix 2 system parameters...
Page 87: ...Appendix 2 System parameters 76...
Page 88: ...Appendix 3 Electromagnetic compliance EMC 77 Appendix 3 Electro magnetic compliance...
Page 93: ...Appendix 3 Electromagnetic compliance EMC 82...
Page 94: ...Appendix 4 The GLASGOW program 83 Appendix 4 The GLASGOW program...
Page 101: ......