S+S Regeltechnik Aerasgard KLG-W Operating Instructions, Mounting & Installation Download Page 7

G  

Installation and Commissioning 

Rev.  2021 - V35

ATTENTION !  

 

The gas inter-exchange in the sensor element happens by diffusion. Depending 
on the changes to the concentration and the flow velocity of the air surrounding 
the sensor, the reaction of the device to the change of concentration may take 
place with a delay. It is essential to choose an installation location for the 
device in which the air stream flows around the sensor. Otherwise the gas 
exchange may be considerably delayed or prevented.  

General information on air quality

There is no universally uniform standard, which could be applied to a 

VOC 

sensor

 (Volatile Organic Compounds). The monitored air contains a large 

number of substances to which the sensor responds, and gas mixtures are 
adjusted. Therefore this sensor does not act selectively, but rather  reflects 
the overall air quality. In principle, one also cannot state precisely what is 
"bad air" or what is "good air" because this is a purely subjective sensation.

Automatic air quality calibration  (permanently active)

We recommend to rinse the sensor every 7 days for 20 minutes or longer with 
fresh air as a basis for automatic calibration. Within a period of approx.  
3 weeks, the minimum output value for air quality is saved. During this period, 
the output signal is standardised to the zero point. The maximum correction 
per interval is limited here. Long-term drifts and the operation-related ageing 
of the sensor  element are thus completely eliminated. 

Every time the device is switched on, the automatic calibration cycle restarts 
and calculates a correction value across the operating period. In case of 
power disruptions of less than 7 days, the calculation algorithm may be  
supported with a manual calibration to adjust the correction value more 
quickly. 

Manual calibration of air quality
Sufficient fresh air must be provided before and after the calibration 
procedure!

We recommend a fresh air supply of at least two hours before starting 
manual calibration.  
The calibration procedure is started by pressing the "ZERO" button  
(for approx. five seconds). This is signalled via the flashing LED. 
Then calibration takes place. During this phase, the LED is continuously 
active and a 60 second countdown runs. 

Putting in operation 

After switching on the device, a self-test and tempering period follows. This 
process takes around 30 - 50 minutes, depending on the ambient conditions. 
Manual calibration can then be performed as an option. Make sure that no 
harmful substances influence the calibration process and that the sensor is 
operated in fresh air.

Switching point setting 

A switching point between 10 % and 95 % of the measuring range can 
be selected using the SET potentiometer. A potential-free changeover 
contact is available as a switch output.

Offset 

The VOC measured value can be adjusted subsequently using the offset  
potentiometer. The adjusting range is ± 10 % of the measuring range.

DIP switches

 

KLQ - W

VOC   sensitivity

DIP 1

DIP 2

LOW

OFF

OFF

MEDIUM  (default)

O N

OFF

HIGH

OFF

O N

IAQ

  (Indoor Air Quality)

O N

O N

Output

DIP 4

Voltage 0 -10 V  (default)

OFF

Current  4...20 mA

O N

Note: 

DIP 3 

is not assigned! 

DIP switches

 

KLQ - SD

VOC   sensitivity

DIP 1

DIP 2

LOW

OFF

OFF

MEDIUM  (default)

O N

OFF

HIGH

OFF

O N

IAQ

  (Indoor Air Quality)

O N

O N

Level IAQ

  (Indoor Air Quality)

 VOC

1

excellent

0

...

19

 

%

no action required

2

good           

20

...

39

 

%

prompt airing recommended

3

moderate       

40

...

59

 

%

airing recommended

4

poor    

60

...

79

 

%

increased airing required

5

unhealthy  

80

...

100

 

%

intense airing necessary

Table according to TVOC guidelines of the German Federal  

Environmental Agency to assess indoor air contamination
(Bundesgesundheitsbl - Gesundheitsforsch -  

Gesundheitsschutz 2007, 50: 990–1005)

Summary of Contents for Aerasgard KLG-W

Page 1: ...uding installation flange self calibrating with multi range switching and active switching output F Notice d instruction Sonde de qualit de l air pour montage en gaine COV resp convertisseur de mesure...

Page 2: ...en mati re plastique Montageschema Mounting diagram Sch ma de montage 250 mm 115 mm 250 mm 115 mm Ma zeichnung Dimensional drawing Plan cot KLQ W KLQ SD KLQ W KLQ SD 14 8 37 8 M16x1 5 32 3 12 13 4 20...

Page 3: ...Ausgang 0 V saubere Luft 10 V verschmutzte Luft KLQ SD 0 10 V fest eingestellt KLQ W 0 10 V oder 4 20 mA B rde 800 ber DIP Schalter w hlbar mit Offset Potentiometer 10 vom Messbereich Relais Ausgang K...

Page 4: ...den um eine schnellere Anpassung des Korrekturwertes zu erreichen Manuelle Kalibrierung der Luftqualit t Vor und w hrend des Kalibriervorganges ist f r ausreichend Frischluft zu sorgen Wir empfehlen e...

Page 5: ...betriebnahme ist diese Anleitung zu lesen und die alle darin gemachten Hinweise sind zu beachten Der Anschluss der Ger te darf nur an Sicherheitskleinspannung und im spannungslosen Zustand erfolgen Um...

Page 6: ...0 10 V or 4 20 mA working resistance 800 selectable via DIP switches with offset potentiometer 10 of the measuring range Relay output KLQ SD without changeover contact KLQ W with potential free change...

Page 7: ...ration to adjust the correction value more quickly Manual calibration of air quality Sufficient fresh air must be provided before and after the calibration procedure We recommend a fresh air supply of...

Page 8: ...onditions In addition the following points are to be observed These instructions must be read before installation and putting in operation and all notes provided therein are to be regarded Devices mus...

Page 9: ...Quality Qualit de l air int rieur Sortie 0 V air propre 10 V air pollu KLQ SD 0 10 V r glage fixe KLQ W 0 10 V ou 4 20 mA charge 800 s lectionnable via interrupteur DIP avec potentiom tre offset 10 de...

Page 10: ...e correction soit adapt e plus rapidement talonnage manuel de la qualit de l air Avant et pendant le processus d talonnage assurer une alimentation en air frais suffisante Nous recommandons une alimen...

Page 11: ...e attentivement la pr sente notice et toutes les consignes qui y sont pr cis es Les raccordements lectriques doivent tre ex cut s HORS TENSION Ne branchez l appareil que sur un r seau de tr s basse te...

Page 12: ...1 A 20 60 1 1 0 3 10 60 C 0 14 1 5 30 RAL 9016 KLQ SD KLQ W 72 x 64 x 37 8 Tyr 1 Tyr 01 M 16 x 1 5 10 4 M12 DIN EN 61076 2 101 PLEUROFORMTM PA6 20 NL 202 5 vmax 30 III EN 60 730 KLQ SD IP54 EN 60 529...

Page 13: ...59 4 60 79 5 80 100 TVOC Bundesgesundheitsbl Gesundheitsforsch Gesundheitsschutz 2007 50 990 1005 DIP KLQ W VO DIP1 DIP2 LOW OFF OFF MEDIUM default ON OFF HIGH OFF ON IAQ Indoor Air Quality ON ON DIP...

Page 14: ...10 0 UB 15 36 UB GND 24 Circuitry Circuitry Output 0V GND Output 0V GND Power supply AC 24V 0V DC 15 36V GND Circuitry Output 0V GND Power supply AC 24V 0V DC 15 36V GND VOC 10 95 0 50 C VOC ZVEI VDE...

Page 15: ...plication potential but do not imply any warranty as to certain product characteristics Since the devices are used under a wide range of different conditions and loads beyond our control their particu...

Page 16: ...Luftg te 0 10V 4 20mA UB GND ffner Schlie er Wechsler 24V 1A 1 2 3 12 11 13 UB 24V AC DC Output air quality 0 10V 4 20mA UB GND Breaker changeover 24V 1A Normally open contact 1 2 3 4 ON DIP SET Zero...

Reviews: