15
The following schedule shall be performed to insure proper operation with the Daily items
performed before each use:
Maintenance Performed
Daily
Weekly
Monthly
Check battery level before each use / Charge battery daily
Wipe Control Box and battery connection with a clean dry rag
Examine hoist for any damage, loose or missing hardware
Test for normal operation
Spray gear assembly with a heavy-duty rust inhibitor/lubricant
such as LPS 3 - Heavy-Duty Inhibitor
Make sure all cable connections are properly secured
Inspect hoist frame, mast, support arm and seat assembly for
rust
Cleansing Performed
– after each use
Rinse seat and seatbelt with fresh water between each use -
Cleanse seat and seatbelt with a disinfectant solution of 1:100
dilution of household bleach to fresh water and then rinse with
fresh water and dry entire hoist daily. In the event of a
contamination incident such patient/user excreta - cleanse seat
and seatbelt immediately with the disinfectant solution*
Cleanse all battery connections with a nylon scouring pad
Cleanse all metallic surfaces with a cleaner wax to maintain the
finish of the hoist
* When using the disinfection solution avoid direct contact with the skin and eyes. In the event of
a contamination incident - immerse the seat belt in the disinfection solution for 10 min. and then
rinse thoroughly with fresh water.
TROUBLE SHOOTING
Be sure the battery is fully charged before troubleshooting.
Hoist
does not rotate
Does hoist raise or lower?
Yes.
1. Check connection to Control Box. Be sure plug is pushed in all the way.
2. Check hand control connection to Control Box for damaged pins.
3. Check connections on terminal block located on frame for loose wires.
4. Check connection cable for damage.
5. Try operating the hoist using the control buttons on the front panel of the controller. If
the hoist rotates when using the buttons on the control box, the hand control is likely the
problem.
6. If the hoist does not rotate when using the buttons on the control box, reverse the
motor cables as follows: Locate the area on the Control Box where the cables are
attached. Swap the actuator cable from connector #2 with the 24v motor cable from
connector #1. Activate the up and down buttons on the hand control. If hoist rotates, the
problem is likely the hand control. If hoist does not rotate, the problem is likely the 24v
motor.
7. Contact your Authorized Reseller or S.R. Smith for replacement information.
Does hoist raise or lower?
No.
1. Check battery charge level.
2. Check battery connection.
3. Use another fully charged battery. If hoist continues to not function, replace the
Control Box.