Documenting Multifunction Calibrator / EN
52
06/04/2018 Version No. 1
Temperature, Thermocouples
(source and measure, 0.1
C & 0.1
F Resolution, Internal Cold Junction
Compensation, thermocouple accuracy not included, 3 minutes after plugging in
thermocouples.)
°C
°F
Range
Accuracy
Range
Accuracy
K
-200 to -150
2.0
-382 to -238
3.6
-150 to 0
1.2
-238 to 32
2.1
0 to 1000
0.8
32 to 1832
1.4
1000 to 1370
1.2
1832 to 2498
2.1
J
-200 to -150
2.0
-382 to -238
3.6
-150 to 0
1.0
-238 to 32
1.8
0 to 1050
0.7
32 to 1922
1.2
E
-200 to -150
1.5
-382 to -238
2.7
-150 to 0
0.9
-238 to 32
1.6
0 to 850
0.7
32 to 1562
1.2
T
-200 to -150
1.5
-382 to -238
2.7
-150 to 0
1.2
-238 to 32
2.1
0 to 400
0.8
32 to 752
1.4
R
0 to 500
1.8
32 to 932
3.2
500 to 1760
1.5
932 to 3200
2.7
S
0 to 500
1.8
32 to 932
3.2
500 to 1760
1.5
932 to 3200
2.7
N
-200 to 0
1.5
-328 to 32
2.7
0 to 1300
0.9
32 to 2372
1.6
L
-200 to 0
0.9
-328 to 32
1.6
0 to 900
0.7
32 to 1652
1.2
U
-200 to 0
1.1
-328 to 32
1.9
0 to 600
0.7
32 to 1112
1.2
B
600 to 800
2.2
1112 to 1472
3.9
800 to 1000
1.8
1472 to 1832
3.2
1000 to 1820
1.4
1832 to 3308
2.5
C
0 to 1800
1.0
32 to 3272
1.8
1800 to 2310
1.5
3272 to 4190
2.7
mV
-10mV to 70mV
0.05mV
-10mV to 70mV
0.05mV