Documenting Multifunction Calibrator / EN
50
06/04/2018 Version No. 1
Voltage Peak to Peak for Sine Wave
(Vpp, 0.3~20KHz, 50% duty cycle, sine wave, 0V offset)
Range(V)
Resolution
Accuracy of Reading
0.1 to 20V
0.001V
5% +/- 0.3V
Voltage Peak to Peak for Non-Sine Wave
(Vpp, 0.3~20KHz, 0V offset)
Range(V)
Resolution
Accuracy of Reading
0.1 to 20V
0.001V
6% +/- 0.4V
Voltage Peak to Peak
(Vpp, 0.3~20KHz, 50% duty cycle, square wave, 0V offset)
Range(V)
Resolution
Accuracy of Reading
1 to 20V
0.001V
6% +/- 0.4V
Voltage of Offset
(Maximum Vpp < 20V)
Range
Resolution
Accuracy of Reading
-5V to 5V
0.001V
5% +/-0.5V +/-5%xVpp
Duty Cycle
(%, square wave, 10 Vpp, 0.3~20KHz)
Range
Resolution
Rise Time of Vpp
Fall Time of Vpp
0 to 100%
1%
10μS max,
5μS typical
15μS max,
7.5μS typical
Pulse
(square wave, 10 Vpp, Offset -5V~+5V)
Range
Resolution
Rise Time
Fall Time
3.0μS to 9999.9μS
0.1
μS
10
μ
S max,
5
μ
S typical
15
μ
S max,
7.5
μ
S typical
10.000mS to 99.999mS
0.001
mS
100.00mS to 999.99mS
0.01
mS