Documenting Multifunction Calibrator / EN
24
06/04/2018 Version No. 1
7. Voltage Input (Measure)
7a. -3V ~ 24V
(1) Turn on the power. Turn sliding switch to V.
(2) Press
button to select MEASURE (input) mode.
(3) Connect Test leads or Alligator clips to MEASURE terminals (red to +,
and black to -).
(4) Then connect the other ends of Test leads or Alligator clips to the object
for measurement.
(5) The calibrator display will show the measurement result.
(6) Data logging function: refer to
“Data logging” chapter.
Remark: 1.
The measurement result is 5-digit (including decimal point and
negative sign).
Warning: 1.
Do not measure over 30V for MEASURE (terminals) to prevent
from damaging the calibrator.
2.
Perform one function at a time and make connection to the specific
terminals only. Remove all other connections to the unused terminals.
Always connect to only one of SOURCE , MEASURE , or TC alone.