HygroPalm HP22 hand-held indicator: 

User Guide 

Instruction Manual 

Document Type 

Page 

9

 of 16 

Document title 

 
 

© 2009-2010; Rotronic AG                                                                                                                             E-M-HP22-V1_21

 

4.3    Internal menu 

Note

: Unauthorized access to the menu can be prevented by disabling the “display menu” setting 

(use the HW4 software > Device Manager > Display) 

 
Main Menu 

Menu Items 

Selections / Information 

Notes 

Device Settings 

 

 

 

 

Units 

Metric / English 

 

 

Back Light 

Key Press / On / Off 

Display backlight  mode 

 

Contrast 

 

LC display contrast adjustment 

 

Trend 

On / Off 

Trend indication on the display 

Device Information 

 

 

 

 

Version 

Firmware version 

 

 

Serial Nbr 

Serial number 

 

 

Address 

RS-485 address 

 

 

Type 

Device type 

 

 

Name 

Device name 

User defined 

Probe Information 

 

 

 

 

Version 

Firmware version 

 

 

Serial Nbr 

Serial number 

 

 

Address 

RS-485 address 

 

 

Name 

Device name 

User defined 

 

SensorTest 

Humidity sensor status 

Off / Good / SQ-Tuned / Bad 

 

Record 

On / Off 

Data recording by the probe 
(max. 2000 values) 

Humidity Adjust 

 

 

 

 

RefValue 

Humidity reference value 

± 0.1 %RH steps 

 

Acquired 

 

Number of cal. points in probe 
memory 

 

<Acquire> 

 

Save cal. point to probe memory 

 

<Delete> 

 

Erases all calibration points 

 

<Adjust> 

 

Effect depends on number of 
calibration points 

Temperature Adjust 

 

 

 

 

RefValue 

Temperature reference value 

± 0.1 ˚C steps

 

 

<Adjust> 

 

1-point adjustment only (offset)

 

 

o

 

Record: 

both the recording mode (start / stop and the log interval cannot be changed from the menu and 

are as configured with the ROTRONIC HW4 software 

o

 

SensorTest: 

Off means that the humidity sensor has not been tested due to the configuration settings of 

the test. For a description of the automatic humidity sensor test and drift compensation (SQ-tuning) see 
documents 

E-T-AC3000-DF-V1

 and 

E-M-HW4v2-F2-006

 

 

Summary of Contents for HygroPalm HP22

Page 1: ...sersdorf Switzerland Document code Unit HygroPalm HP22 hand held indicator User Guide Instruction Manual Document Type Page 1 of 16 Document title 2009 2010 Rotronic AG E M HP22 V1_21 HygroPalm HP22 H...

Page 2: ...ings 5 3 3 Interaction between the HP22 and probe functions 7 4 Operation 8 4 1 Display 8 4 2 Keypad 8 4 3 Internal menu 9 4 4 Changing the parameters shown on the display 10 4 5 Changing the unit sys...

Page 3: ...nd 100 to 200 C 148 to 392 F The temperature operating range of the HP22 electronics is limited to 10 60 C Based on the ROTRONIC AirChip 3000 digital technology the HP22 indicator and HygroClip 2 prob...

Page 4: ...23 hand held calibrator This allows showing the data measured by the HP22 on the other instrument display and also allows using the other instrument internal menu to do for example a humidity and temp...

Page 5: ...d data can be downloaded to a PC with the HW4 software time stamped and viewed MEASUREMENT LOOP VALIDATION Functions Description Simulator mode Used to make the HP22 generate fixed values for the humi...

Page 6: ...djustment K Device write protection Disabled HP22 and probe Menu access from keypad Enabled Limit humidity output to 100 RH Enabled probe Out of limit value digital display alarm Disabled HP22 and pro...

Page 7: ...e name of the HC2 probe is not displayed by HW4 and is replaced with the HP22 Input Name Calculation X X Psychrometric calculation HP22 setting overrides HC2 probe setting Data refresh rate X When set...

Page 8: ...idity or dew frost point and the bottom line corresponds to temperature The display can be configured to show a trend indicator on each line increasing value decreasing value In the event of an alarm...

Page 9: ...irmware version Serial Nbr Serial number Address RS 485 address Name Device name User defined SensorTest Humidity sensor status Off Good SQ Tuned Bad Record On Off Data recording by the probe max 2000...

Page 10: ...dvice for measuring humidity The most common source of error when measuring relative humidity is a difference between the temperature of the probe and the temperature of the environment At a humidity...

Page 11: ...or to using this cable the ROTRONIC USB driver must be installed on the PC Both the driver and the installation instructions document E M HW4v2 Main are located on the HW4 CD As an alternative cable A...

Page 12: ...done prior to adjusting humidity o The calibration point is automatically deleted from the probe memory after an adjustment o Because the HP22 has no real time clock the date of the adjustment is not...

Page 13: ...roups Search for USB Masters After finding the HP22 with HW4 expand the device tree to see the HP22 functions Select Probe and Probe Adjustment For further instructions see HW4 manual E M HW4v2 A2 001...

Page 14: ...xing ratio by weight R Vapor concentration at saturation Dvs Vapor partial pressure E Vapor saturation pressure Ew Start up time and data refresh rate HP22 Start up time 3 s typical Data refresh rate...

Page 15: ...s the HW4 configuration software service cables calibration accessories and spare dust filters please see document E M HC2 accessories 9 Supporting documents Document File Name Contents E M HC2 Probes...

Page 16: ...6 Document title 2009 2010 Rotronic AG E M HP22 V1_21 E T HumiDefs Humidity Definitions Note All document file names have an extension corresponding to the document release number This extension is no...