5.4    Periodic calibration check of the probe 

 
Both the Pt 100 RTD temperature sensor used in the probe and associated electronics are very stable 
and should not require any calibration after the initial factory adjustment. Long term stability of the 
ROTRONIC Hygromer humidity sensor is typically better than 1 %RH per year. For maximum accuracy, 
calibration of the probe should be verified every 6 to 12 months. Applications where the probe is 
exposed to significant pollution may require more frequent verifications. 
 

a) Procedure for adjusting the probe from the HP22 keypad 
 

Temperature adjustment 

 
The keypad of the HP22 allows only a 1-point adjustment of temperature. The effect of a 1-point 
temperature adjustment is to add the same offset to all measured values. 
  

  When temperature is stable, press the MENU key to show the internal menu on the display 
  Use the (-) key to select Temperature Adjust and press the ENTER key 
  Make sure that the text line beginning with RefValue is highlighted and press the ENTER key 
  Use the (+) or (-) key to change the reference value to match the temperature reference 
  Use the (-) key to highlight the Adjust text line and press the ENTER key 
  The HP22 

confirms the adjustment with the message “Adjust OK”  

  Press the MENU key twice to exit the menu and return the HP22 to normal operation 

 

Notes: 

 

o

 

Any temperature adjustment should be done prior to adjusting humidity  

o

 

The calibration point is automatically deleted from the probe memory after an adjustment 

o

 

Because the HP22 has no real time clock, the date of the adjustment is not written to the probe. If 
retaining the adjustment date is important, use the HW4 software to adjust the probe. 

 
Humidity adjustment 

 
The keypad of the HP22 allows a multi-point adjustment of humidity. The effect of a humidity adjustment 
depends on the number of calibration points present in the probe memory prior to the adjustment: 
 

-

  one calibration point: offset  added to all measured values 

-

  two calibration points: offset and slope 

-

  three or more calibration points: offset, slope, linearity 

 
For maximum accuracy use at least 3 to 4 calibration points distributed equally across the measurement 
range of interest. The calibration points (maximum 100) can be acquired in any order but we recommend 
going from low humidity values to high humidity values. 
 

  When humidity is stable, press the MENU key to show the internal menu on the display 
  Use the (-) key to select Humidity Adjust and press the ENTER key 
  Make sure that the text line beginning with RefValue is highlighted and press the ENTER key 
  Use the (+) or (-) key to change the reference value to match the reference humidity 
  Use the (-) key to highlight the Acquire text line and press the ENTER key 
 

Note that the “Acquired” counter is incremented by 1 (number of calibration points in the probe 
memory) 

  When all calibration points have been acquired, use the (-) key to highlight the Adjust text line and 

press the ENTER key. 

Do not adjust the probe before having acquired all calibration points

  The HP22 

confirms the adjustment with the message “Adjust OK”  

  Press the MENU key twice to exit the menu and return the HP22 to normal operation 

Summary of Contents for HygroPalm HP22

Page 1: ...sersdorf Switzerland Document code Unit HygroPalm HP22 hand held indicator User Guide Instruction Manual Document Type Page 1 of 16 Document title 2009 2010 Rotronic AG E M HP22 V1_21 HygroPalm HP22 H...

Page 2: ...ings 5 3 3 Interaction between the HP22 and probe functions 7 4 Operation 8 4 1 Display 8 4 2 Keypad 8 4 3 Internal menu 9 4 4 Changing the parameters shown on the display 10 4 5 Changing the unit sys...

Page 3: ...nd 100 to 200 C 148 to 392 F The temperature operating range of the HP22 electronics is limited to 10 60 C Based on the ROTRONIC AirChip 3000 digital technology the HP22 indicator and HygroClip 2 prob...

Page 4: ...23 hand held calibrator This allows showing the data measured by the HP22 on the other instrument display and also allows using the other instrument internal menu to do for example a humidity and temp...

Page 5: ...d data can be downloaded to a PC with the HW4 software time stamped and viewed MEASUREMENT LOOP VALIDATION Functions Description Simulator mode Used to make the HP22 generate fixed values for the humi...

Page 6: ...djustment K Device write protection Disabled HP22 and probe Menu access from keypad Enabled Limit humidity output to 100 RH Enabled probe Out of limit value digital display alarm Disabled HP22 and pro...

Page 7: ...e name of the HC2 probe is not displayed by HW4 and is replaced with the HP22 Input Name Calculation X X Psychrometric calculation HP22 setting overrides HC2 probe setting Data refresh rate X When set...

Page 8: ...idity or dew frost point and the bottom line corresponds to temperature The display can be configured to show a trend indicator on each line increasing value decreasing value In the event of an alarm...

Page 9: ...irmware version Serial Nbr Serial number Address RS 485 address Name Device name User defined SensorTest Humidity sensor status Off Good SQ Tuned Bad Record On Off Data recording by the probe max 2000...

Page 10: ...dvice for measuring humidity The most common source of error when measuring relative humidity is a difference between the temperature of the probe and the temperature of the environment At a humidity...

Page 11: ...or to using this cable the ROTRONIC USB driver must be installed on the PC Both the driver and the installation instructions document E M HW4v2 Main are located on the HW4 CD As an alternative cable A...

Page 12: ...done prior to adjusting humidity o The calibration point is automatically deleted from the probe memory after an adjustment o Because the HP22 has no real time clock the date of the adjustment is not...

Page 13: ...roups Search for USB Masters After finding the HP22 with HW4 expand the device tree to see the HP22 functions Select Probe and Probe Adjustment For further instructions see HW4 manual E M HW4v2 A2 001...

Page 14: ...xing ratio by weight R Vapor concentration at saturation Dvs Vapor partial pressure E Vapor saturation pressure Ew Start up time and data refresh rate HP22 Start up time 3 s typical Data refresh rate...

Page 15: ...s the HW4 configuration software service cables calibration accessories and spare dust filters please see document E M HC2 accessories 9 Supporting documents Document File Name Contents E M HC2 Probes...

Page 16: ...6 Document title 2009 2010 Rotronic AG E M HP22 V1_21 E T HumiDefs Humidity Definitions Note All document file names have an extension corresponding to the document release number This extension is no...

Reviews: