background image

User Manual 

 

75 

  

Calibration Kit

 

Industrial Version 

74CK011-170 

 

 

Rosenberger Hochfrequenztechnik GmbH & Co. KG  
P.O.Box 1260       D-84526 Tittmoning       Germany 
www.rosenberger.com

 

Tel.    : +49 8684 18-0  
 
Email : [email protected]

 

Page 

 

12 / 15

 

 

D

ie

s

e

s

 D

o

k

u

m

e

n

is

u

rh

e

b

e

rr

e

c

h

tl

ic

h

 g

e

s

c

h

ü

tz

 T

h

is

 d

o

c

u

m

e

n

is

 p

ro

te

c

te

d

 b

y

 c

o

p

y

ri

g

h

 R

o

s

e

n

b

e

rg

e

H

o

c

h

fr

e

q

u

e

n

z

te

c

h

n

ik

 G

m

b

H

 &

 C

o

K

G

 

6.1.3 Cleaning procedure 

 
Items required: 
 

  Low pressure compressed air (solvent free). 

  Cotton swabs. 

  Lint free cleaning cloth. 

  Isopropanol. 

  Illuminated magnifier or microscope (factor 5 to 10) 

 

NOTE:  Isopropanol that contains additives should not be used for cleaning connectors as 

it may cause damage to plastic dielectric support beads in coaxial and microwave 
connectors.  It  is  important  to  take  all  necessary  safety  precautions  when  using 
chemicals or solvents. 

 

  Remove  loose  particles  on  the  mating  surfaces  and  threads  etc.  using  low-pressure 

compressed air. 

  Clean surfaces using Isopropanol on cotton swabs or lint free cloth. Use only sufficient solvent 

to  clean  the  surface.  When  using  swabs  or  lint  free  cloth,  use  the  least  possible  pressure  to 
avoid damaging connector surfaces. Do not spray solvents directly on to the connector surfaces 
or use contaminated solvents. 

  Use the low-pressure compressed air once again to remove any remaining small particles and 

to  dry  the  surfaces  of  the  connector  to  complete  the  cleaning  process  before  using  the 
connector. 

 

6.1.3 Cleaning connectors on static sensitive devices 

 
Special  care  is  required  when  cleaning  connectors  on  test  equipment  containing  electrostatic 
sensitive devices. When cleaning such connectors always wear at least a grounded wrist strap and 
observe correct procedures.  

 

6.2 Interface Dimensions and Gauging 

 
It  is  of  the  utmost  importance  that  connectors  do  not  damage  the  test  equipment  interfaces  to 
which  they  are  connected.  Poor  performance  of many  coaxial  devices  and  cable  assemblies  can 
often  be traced to  poor construction  and  non-compliance  with the mechanical  specifications.  The 
mechanical  gauging  of  connectors  is  essential  to  ensure  correct  fit  and  to  achieve  the  best 
performance.  This  means  that  all  coaxial  connectors  fitted  on  all  equipment,  cables  and 
terminations  etc.  should  be  gauged  on  a  regular  basis  in  order  to  detect  any  out  of  tolerance 
conditions that may impair the electrical performance. 
 

6.2.1 Gauging connectors 

 
A connector should be gauged before it is used for the very first time or if someone else has used 
the device to which it is connected. 

Summary of Contents for 74CK011-170

Page 1: ...hnik GmbH Co KG P O Box 1260 D 84526 Tittmoning Germany www rosenberger com Tel 49 8684 18 0 Email info rosenberger com Page 1 15 Dieses Dokument ist urheberrechtlich gesch tzt This document is protec...

Page 2: ...4 Calibration Certificate 1 5 2 Specifications Environmental Requirements 2 1 Electrical and mechanical Specifications 2 2 3 Standard Definitions Standard Definitions 3 1 Installation of Standard Defi...

Page 3: ...his calibration kit allow you to perform one port OSL OSM or two port OSLT TOSM and TRM TMR calibrations from DC up to 6 GHz for measurements of Rosenberger components with F 75 connectors Components...

Page 4: ...a 74K12L 001S3 1 Short circuit plug 74S12S 001S3 1 Short circuit jacka 74K12S 001S3 1 Calibration load plug 74S150 C11S3 1 Calibration load jacka 74K150 C11S3 1 Calibration Adaptor plug plug 74S121 S2...

Page 5: ...PNA Keysight Agilent and ZVA Rohde Schwarz Calibration Certificate as PDF file Standard Definitions Cards Printed Standard Definitions that can be used on nearly all Vector Network Analyzers Kit Info...

Page 6: ...ure range for which the specifications are valid is defined in the specific data sheets 2 2 Electrical and mechanical Specifications Electrical and mechanical specifications are defined for all calibr...

Page 7: ...a more complex process Please consult the VNA operating manual Some network analyzers allow a step by step input of the model based standard definitions during the VNA calibration without previously l...

Page 8: ...thmic Sweep range start frequency stop frequency Number of points frequency step size Source power IF bandwidth small IF bandwidth basically increases the dynamic range Sweep time Connect test port ca...

Page 9: ...ilable In cases when the two VNA test ports have different connector systems installed e g RPC N 75 and F 75 UOSM can handle this situation A second calibration kit is needed Calibration Procedure Ada...

Page 10: ...cal Rosenberger sales office Please use the Service Form at the end of this manual if there are any known problems with your calibration kit 5 4 Extent of Rosenberger Re Calibration Services Every cal...

Page 11: ...s will usually also have raised edges around them and will make less than perfect contact further to this raised edges on mating interfaces will make dents in other connectors to which they are mated...

Page 12: ...olvents directly on to the connector surfaces or use contaminated solvents Use the low pressure compressed air once again to remove any remaining small particles and to dry the surfaces of the connect...

Page 13: ...grity non destructive mating and electrical performance of the connector The mechanical gauging of coaxial connectors will detect and prevent the following problems Inner conductor protrusion This may...

Page 14: ...ways wear a grounded wrist strap and work on a grounded conductive table mat 2 Visually inspect the connectors 3 If necessary clean the connectors Refer to 4 1 Connector Care 4 Check connector gage ac...

Page 15: ...onnecting a sliding load leave the centre conductor pullback handle in the locked position Use the combi wrench to prevent the device body from turning Use torque wrench to loosen the connecting nut C...

Reviews: