5. Calibration Kit Service 

5.1 Calibration interval 

 
Rosenberger suggests a calibration interval of 12 months. Depending on the frequency of use, the 
wear and the actual requirements a different calibration interval might be suitable. This lies in the 
responsibility of the user.  
 

5.2 Spare parts 

 
Each  part  of  this  kit  can  be  ordered  separately  as  spare  part  with  reference  to  the  part  number 
shown  in  Table  1-1  and  the  quantity  desired.  Telephone  or  send  your  order  to  your  local 
Rosenberger sales office. 
 

5.3 Where to send a Kit for Re-Calibration / Repair 

 
In  case  of  re-calibration  or  repair  contact  your  local  Rosenberger  sales  office.  Please  use  the 
Service Form at the end of this manual if there are any known problems with your calibration kit 
 

5.4 Extent of Rosenberger Re-Calibration Services 

 
Every calibration kit sent for re-calibration or repair the following services are performed 

  Cleaning of connector interfaces 

  Optical check of connector mating planes, contact pins and contact fingers 

  Check of connector gauge for impermissible protrusion and significant recession 

  Electrical measurement of all calibration standards 

  Issuing of calibration certificates including measurement values and uncertainties 

  Individual  optimizing  of capacitive  coefficients  C0  to  C3 for  all  Open  calibration  standards  and 

short  inductance  coefficients  L0  to  L3  for  all  Short  calibration  standards  for  best  fitted  model 
based standard definitions. 

 
In case of any defect or abnormality a cost estimation will be issued before continuing repair work. 
 
If a calibration standard has to be replaced or modified, the incoming status is additionally reported 
in  the  calibration  certificate.  This  service  is  excluded  when  this  measurement  may  damage  the 
Rosenberger measurement setup. 
 
 
 

 

 

 

Summary of Contents for 74CK011-170

Page 1: ...hnik GmbH Co KG P O Box 1260 D 84526 Tittmoning Germany www rosenberger com Tel 49 8684 18 0 Email info rosenberger com Page 1 15 Dieses Dokument ist urheberrechtlich gesch tzt This document is protec...

Page 2: ...4 Calibration Certificate 1 5 2 Specifications Environmental Requirements 2 1 Electrical and mechanical Specifications 2 2 3 Standard Definitions Standard Definitions 3 1 Installation of Standard Defi...

Page 3: ...his calibration kit allow you to perform one port OSL OSM or two port OSLT TOSM and TRM TMR calibrations from DC up to 6 GHz for measurements of Rosenberger components with F 75 connectors Components...

Page 4: ...a 74K12L 001S3 1 Short circuit plug 74S12S 001S3 1 Short circuit jacka 74K12S 001S3 1 Calibration load plug 74S150 C11S3 1 Calibration load jacka 74K150 C11S3 1 Calibration Adaptor plug plug 74S121 S2...

Page 5: ...PNA Keysight Agilent and ZVA Rohde Schwarz Calibration Certificate as PDF file Standard Definitions Cards Printed Standard Definitions that can be used on nearly all Vector Network Analyzers Kit Info...

Page 6: ...ure range for which the specifications are valid is defined in the specific data sheets 2 2 Electrical and mechanical Specifications Electrical and mechanical specifications are defined for all calibr...

Page 7: ...a more complex process Please consult the VNA operating manual Some network analyzers allow a step by step input of the model based standard definitions during the VNA calibration without previously l...

Page 8: ...thmic Sweep range start frequency stop frequency Number of points frequency step size Source power IF bandwidth small IF bandwidth basically increases the dynamic range Sweep time Connect test port ca...

Page 9: ...ilable In cases when the two VNA test ports have different connector systems installed e g RPC N 75 and F 75 UOSM can handle this situation A second calibration kit is needed Calibration Procedure Ada...

Page 10: ...cal Rosenberger sales office Please use the Service Form at the end of this manual if there are any known problems with your calibration kit 5 4 Extent of Rosenberger Re Calibration Services Every cal...

Page 11: ...s will usually also have raised edges around them and will make less than perfect contact further to this raised edges on mating interfaces will make dents in other connectors to which they are mated...

Page 12: ...olvents directly on to the connector surfaces or use contaminated solvents Use the low pressure compressed air once again to remove any remaining small particles and to dry the surfaces of the connect...

Page 13: ...grity non destructive mating and electrical performance of the connector The mechanical gauging of coaxial connectors will detect and prevent the following problems Inner conductor protrusion This may...

Page 14: ...ways wear a grounded wrist strap and work on a grounded conductive table mat 2 Visually inspect the connectors 3 If necessary clean the connectors Refer to 4 1 Connector Care 4 Check connector gage ac...

Page 15: ...onnecting a sliding load leave the centre conductor pullback handle in the locked position Use the combi wrench to prevent the device body from turning Use torque wrench to loosen the connecting nut C...

Reviews: