4. VNA Calibration 

 

4.1 VNA Calibration 

 

A  VNA  is  only  as  useful  as  the  accuracy  of  the  measurements  it  makes,  and  this  requires  the 
instrument  to  be  calibrated.  The  calibration  process  employs  a  technique  called  vector  error 
correction,  in  which  error  terms  are  characterized  using  known  standards  so  that  errors  can  be 
removed from actual measurements. The process of removing these errors requires the errors and 
measured quantities to be measured vectorially (thus the need for a vector network analyzer). 
 
 

Main setting parameters

 
It is good practice to start your measurement by selecting the main setting parameters. 
 

  Sweep type (linear / logarithmic) 

  Sweep range (start frequency / stop frequency) 

  Number of points (frequency step size) 

  Source power 

  IF bandwidth (small IF bandwidth basically increases the dynamic range) 

  Sweep time 

 
Connect test port cables and adapters to your VNA as needed for your specific measurement task. 
 

Calibration techniques 

 
A  very  large  number  of  calibration  techniques  have  been  developed  to  calibrate  a  VNA.  The 
following list describes calibration techniques you can apply with this calibration kit along with their 
features and applicability for use in specific situations.  
 

One-port calibration 

 

OSL (OSM)

   

 

Open-Short-Load (Open-Short-Match) 

 
When  only  reflection  measurements  are  needed,  the  fastest  and  easiest  calibration 
technique is OSL. It requires 3 connections of calibration standards on the reference plane. 

 

Two-port calibrations 

 

OSLT (TOSM)

 

 

Open-Short-Load-Thru (Thru-Open-Short-Match) 

 
OSLT  is  the  most  commonly  used  two-port  calibration  method.  The  Thru  connections  are 
performed by using well-known, precision calibration adapters in male-male or female-female 
configuration  or  as  a  direct  connection  of  the  male  and  the  female  test  ports.  
Note: This Calibration Kit does not include in-series calibration adapters. 
 
 

Summary of Contents for 74CK011-170

Page 1: ...hnik GmbH Co KG P O Box 1260 D 84526 Tittmoning Germany www rosenberger com Tel 49 8684 18 0 Email info rosenberger com Page 1 15 Dieses Dokument ist urheberrechtlich gesch tzt This document is protec...

Page 2: ...4 Calibration Certificate 1 5 2 Specifications Environmental Requirements 2 1 Electrical and mechanical Specifications 2 2 3 Standard Definitions Standard Definitions 3 1 Installation of Standard Defi...

Page 3: ...his calibration kit allow you to perform one port OSL OSM or two port OSLT TOSM and TRM TMR calibrations from DC up to 6 GHz for measurements of Rosenberger components with F 75 connectors Components...

Page 4: ...a 74K12L 001S3 1 Short circuit plug 74S12S 001S3 1 Short circuit jacka 74K12S 001S3 1 Calibration load plug 74S150 C11S3 1 Calibration load jacka 74K150 C11S3 1 Calibration Adaptor plug plug 74S121 S2...

Page 5: ...PNA Keysight Agilent and ZVA Rohde Schwarz Calibration Certificate as PDF file Standard Definitions Cards Printed Standard Definitions that can be used on nearly all Vector Network Analyzers Kit Info...

Page 6: ...ure range for which the specifications are valid is defined in the specific data sheets 2 2 Electrical and mechanical Specifications Electrical and mechanical specifications are defined for all calibr...

Page 7: ...a more complex process Please consult the VNA operating manual Some network analyzers allow a step by step input of the model based standard definitions during the VNA calibration without previously l...

Page 8: ...thmic Sweep range start frequency stop frequency Number of points frequency step size Source power IF bandwidth small IF bandwidth basically increases the dynamic range Sweep time Connect test port ca...

Page 9: ...ilable In cases when the two VNA test ports have different connector systems installed e g RPC N 75 and F 75 UOSM can handle this situation A second calibration kit is needed Calibration Procedure Ada...

Page 10: ...cal Rosenberger sales office Please use the Service Form at the end of this manual if there are any known problems with your calibration kit 5 4 Extent of Rosenberger Re Calibration Services Every cal...

Page 11: ...s will usually also have raised edges around them and will make less than perfect contact further to this raised edges on mating interfaces will make dents in other connectors to which they are mated...

Page 12: ...olvents directly on to the connector surfaces or use contaminated solvents Use the low pressure compressed air once again to remove any remaining small particles and to dry the surfaces of the connect...

Page 13: ...grity non destructive mating and electrical performance of the connector The mechanical gauging of coaxial connectors will detect and prevent the following problems Inner conductor protrusion This may...

Page 14: ...ways wear a grounded wrist strap and work on a grounded conductive table mat 2 Visually inspect the connectors 3 If necessary clean the connectors Refer to 4 1 Connector Care 4 Check connector gage ac...