Rosemount Model P300 Pressure Validator
3-4
Setting Main Display Mode
FIGURE 3-2. Typical Main Display in
Voltage Mode.
Set Pressure as Main Display
Press the
P
key once. A beep will sound and the display will change.
Set Current as Main Display
Press the
I
key once. A beep will sound and the display will change.
Set Voltage as Main Display
Press the
V
key once. A beep will sound and the display will change.
NOTE
MAIN
DISPLAY
:
PRESSURE
is the default display mode when the validator
is shipped from the manufacturer.
Setting Dual Display Mode
FIGURE 3-3. A Typical Dual Display
with Pressure in Main Display and
Voltage in Secondary Display.
Set Pressure as Main Display with Voltage as Secondary Display
Press the
V
and
P
keys together. A beep will sound and the display will
change.
Set Pressure as Main Display with Current as Secondary Display
Press the
I
and
P
keys together. A beep will sound and the display will
change.
Zeroing the Display
The
ZERO
key is used to zero displays in all calibration modes. Zeroing
should be done before all calibrations to ensure correct readings.
To zero a display
Press the desired function key and
ZERO
simultaneously. A beep sounds
and the secondary display reads
ZEROING
__
.
If the instrument is in a
different mode than the one being zeroed, the secondary display reads
INCORRECT
MODE
.
P
300
_0
7A
P
300
_08
A
Summary of Contents for P300
Page 2: ......
Page 4: ...Rosemount Model P300 Pressure Validator ii ...
Page 14: ...Rosemount Model P300 Pressure Validator 2 6 ...
Page 20: ...Rosemount Model P300 Pressure Validator 3 6 ...
Page 30: ...Rosemount Model P300 Pressure Validator 4 10 ...
Page 34: ...Rosemount Model P300 Pressure Validator 5 4 ...
Page 36: ...Rosemount Model P300 Pressure Validator 6 2 ...
Page 45: ...I 3 Index ...
Page 46: ...Rosemount Model P300 Pressure Validator I 4 ...
Page 47: ... 1 ...