background image

7

PACKAGE CONTENTS

1.  DT6030 TOP TENS™ Device

2.  1 pack of 4 Electrodes

3.  3 AAA batteries

4.  2 lead wires

5.  Instruction Manual.

6.  Quick Start Guide

TOP TENS™ device

       

4pcs of Electrodes

          

2 lead wires

     

3 of AAA batteries

    

Instruction Manual

ABOUT THE TOP TENS™

1.  On / + key: Power on / adjust / increase setting key

2.  Off / - key: Power off / adjust / decrease setting key

3.  Mode key: Therapy Mode (Program) / Timer selection Key

4.  CH key: select CH1 / CH2

5.  Connection Socket CH1

6.  Connection Socket CH2

7.  Batter compartment

8.  Battery cover

The LCD Display:

1.  Intensity level for CH1
2.  Intensity level for CH2
3.  Low battery indicator
4.  Program number
5.  Timer 

Summary of Contents for TOP TENS DT6030

Page 1: ...TOP TENS DT6030 Instruction Manual Please keep this instruction manual safe for future use...

Page 2: ...TMENT 8 Preparing the Skin for Treatment 8 Inserting 3 AAA Batteries 8 Connecting the Lead Wires to the TOP TENS 9 Turning ON the TOP TENS 9 Setting Treatment Program Time and Intensity Level 9 Turnin...

Page 3: ...timulation to carotid sinus region of the neck Unknown etiology Do not use the TOP TENS if pain symptoms are undiagnosed Use only after the origin cause of pain has been determined by your physician H...

Page 4: ...gion or on the head because of the risk of stimulating inappropriate muscles and organs Do not place the electrodes over any recent scars broken or inflamed areas of infection or susceptibility to acn...

Page 5: ...sarean section you must consult your physician before use The heavy days of your period have finished because vigorous abdominal exercise is not recommended during this time Precautions Read User Manu...

Page 6: ...s moistening skin prior to treatment improves treatment efficacy use of cold packs on treated muscles after treatment is also recommended The TOP TENS should only be used with the leads electrodes and...

Page 7: ...s not disappear after 30 minutes from the removal of electrodes do not use the stimulator again until after the excessive redness has disappeared Turn off the TOP TENS if the stimulation feels unpleas...

Page 8: ...he electrodes firmly to the skin Do not use electrodes which do not adhere properly to the skin If your skin is red under the electrodes after a session do not start another session in the same area u...

Page 9: ...Instruction Manual ABOUT THE TOP TENS 1 On key Power on adjust increase setting key 2 Off key Power off adjust decrease setting key 3 Mode key Therapy Mode Program Timer selection Key 4 CH key select...

Page 10: ...d extend the life of your electrodes 5 When removing electrodes always remove by pulling in the direction of hair growth 6 It may be helpful to apply skin lotion on electrode placement area when not w...

Page 11: ...will flash on the screen Press the On increase or Off decrease button to scroll through treatment programs P1 thru P8 see page 14 for treatment program details Press the Mode button to lock in the se...

Page 12: ...power off the device If any part of the device has a setting that is blinking on the screen you need to lock in the setting by pressing either the Mode or CH buttons before the device will allow you t...

Page 13: ...restart unless you reset the therapy time Automatic Shut off If no button is pushed for 60 seconds when in programming mode the TOP TENS will automatically turn off The TOP TENS automatically turns of...

Page 14: ...TENS and electrodes Do not pull on the lead wires but on the connectors attached to the ends of the lead wires Store the TOP TENS with the lead wires in a clean dry place Electrodes The electrodes are...

Page 15: ...r if it is not possible to remedy in the manner described Defect Cause Remedy The TOP TENS does not turn on No battery or bad battery Replace battery The TOP TENS turns on and then off again Battery n...

Page 16: ...Symmetrical Bi Phasic square pulse Power Source 3 x AAA 4 5 Volt batteries All electrical specifications are 20 at 500 load PROGRAM DETAILS Program Intensity Max Pulse Width Pulse Rate Mode P1 100 mA...

Page 17: ...uctions accidents abuse alterations or disassembly by unauthorized individuals The warranty applies to the main device and necessary parts and labor relating thereto Batteries lead wires electrodes an...

Page 18: ...actured for Current Solutions a Roscoe Medical Company 21973 Commerce Parkway Strongsville OH 44149 1 800 871 7858 www currentsolutionsnow com www roscoemedical com Model No WL 2407 IM 24 036 RevA12 F...

Reviews: