Remote Control Commands
R&S
®
TS7124AS
53
Instructions Handbook 1179.2948.02 ─ 01
The expected time is defined by the
command.
To query the state at any time, use the
Usage:
Event
TIMEOUT:<seconds>
Sets the timeout value for a door status error.
Per default, the door is assumed to open within 4
seconds and to close within 4
sec-
onds. However, you can adjust the
by two control screws on the rear side
of the chamber. If you increase or decrease the opening or closing speed, you must
adjust the timeout setting accordingly, to avoid delayed or false door blockage detec-
tion (
.
Set the timeout value to the actual physical opening or closing time, whichever takes
longer.
Parameters:
<seconds>
Number of seconds that the door control system allows for the
opening time and for the closing time of the door, before it
assumes a mechanical failure. If the successful completion of
opening or closing takes longer than the specified timeout, the
control system returns a
DOOR ERR\r
message and
the pneumatic system.
Range:
3 to 8, unit = s, default = 4
Usage:
Setting only
DOOR_TOUCH_CLOSE:<boolean>
Enables or disables a function for pushing the door to close it, as described in
ter 7.3.2, "Push-Button Door Operation"
default, the function is enabled. See also
Chapter 10, "Troubleshooting and Repair"
Parameters:
<boolean>
0
Disables the function.
1
Enables the function.
*RST:
1
Usage:
Setting only