Usage
R&S
®
TS-PSAM
7
User Manual 1142.9878.12 ─ 12
1.2.2.2
ICT Measurements with R&S TS-PICT
On this topic, see also
and
Further in-circuit measurements can be made in conjunction with the R&S TS-PICT
module (ICT expansion module). These are:
●
Diode and transistor test
●
Guarded resistance measurements (3, 4 and 6 wires)
●
Impedance measurements (3, 4 and 6 wires)
For this purpose, the R&S TS-PICT module provides a special AC voltage source
(AOS) and current measurement unit (CMU).
The UUT is connected for the in-circuit test (ICT) via the R&S TS-PMB module (matrix
module B).
A
B2
R
x
A1
=
R1
R2
A2
C
D
B1
CMU
DCS
Figure 1-5: Cross-connection in a guarded resistance measurement (6-wire)
Features of the R&S TS-PSAM