Working with the Leveling Tool
Software Utility
27
Getting Started 1331.3069.02 ─ 03
The manual calibration steps must be performed in ascending order. Requesting man-
ual calibration step M causes the Leveling Tool to execute missing steps 1 to M-1 auto-
matically, preserving the results of steps < M that have already been executed and
dropping the results of steps > M.
Automatic calibration always executes all applicable steps.
Automatic Calibration > Calibrate All
.............................................................................27
Manual Calibration > 1. Calibration: Reference Receiver
............................................. 27
Manual Calibration > 2. Calibration: RF Input Power
....................................................28
Manual Calibration > 3. Calibration: Waveguide Attenuator
......................................... 29
Automatic Calibration > Calibrate All
Performs calibration steps
Manual Calibration > 1. Calibration: Reference Receiver
Performs a reference receiver (a wave) calibration for the related frequency converter
port and activates it in the analyzer firmware.
Figure 9-2: Reference receiver calibration (R&S
ZNA firmware)
Calibration Commands