Working with the Leveling Tool
Software Utility
32
Getting Started 1331.3069.02 ─ 03
Loading the calibration data manually
You can also load previously generated calibration data manually using the analyzer's
graphical user interface. This involves two steps:
●
R&S
ZNA
:
–
Open the "Leveling Dataset Configuration" dialog (System – [Setup] > "Fre-
quency Converter" > "Leveling Dataset"; see
).
Select "Power Control" method "User" for the respective converter port, and
click the ellipsis button in the "User Control File Directory" to browse for the lev-
eling dataset.
–
Open the "Calibration Manager" dialog (Channel > [Cal] > "Use Cal" > "Cal
Manager...").
Select the recorded receiver calibration from the "Calibration Pool" and apply it
to the active channel (see
●
R&S
ZVA or R&S
ZVT:
–
Open the "Frequency Converter" tab of the "System Configuration" dialog (see
). Click the ellipsis button in the "Power Control" column of the "Port
Config" table to open the Power Control dialog (see
). Select "RF In
Power (and Electronic Attenuator) with Data Set" and click "Read User Data
Set" to load the calibration data.
–
Open the "Calibration Manager" dialog ("Channel > Calibration > Calibration
Manager..."). Select the recorded receiver calibration from the "Calibration
Pool" and apply it to the active channel (see
).
Operating the Leveling Tool