Working with the Leveling Tool
Software Utility
31
Getting Started 1331.3069.02 ─ 03
In the
R&S
ZVA or R&S
ZVT
firmware, this selects control method "RF In Power (and
Electronic Attenuator) with Data Set" for the related frequency converter port and acti-
vates the "Data Set" created by the Leveling Tool.
Figure 9-5: Power Control configuration
Here, the user data sets are stored in a child directory of
C:\Rohde&Schwarz\NWA\FrequencyConverter\
. The directory name is automat-
ically generated using the frequency converter type and the current date and time.
9.4
Operating the Leveling Tool
Operation of the Leveling Tool is straightforward. For each frequency converter "Port i "
that requires output power leveling, connect the power sensor to the waveguide test
port, activate the corresponding tab in the main window of the Leveling Tool and pro-
ceed as follows:
1. Select
Automatic Calibration > Calibrate All
2. To apply the recorded values, click "Activate Cal Data" in the "Finish" panel of the
main window.
Operating the Leveling Tool