Command Reference
R&S
®
CMW-KM645
43
User Manual 1173.9740.02 ─ 04
Example:
See
Continuous and AF Measurements
Firmware/Software:
V1.0.15.20
Manual operation:
See
CONFigure:FMSTereo:MEAS<i>:SCONdition
<StopCondition>
Qualifies whether the measurement is stopped after a failed limit check or continued.
SLFail
means that the measurement is stopped and reaches the
RDY
state as soon
as one of the results exceeds the limits.
Parameters:
<StopCondition>
NONE | SLFail
NONE:
Continue measurement irrespective of the limit check
SLFail:
Stop measurement on limit failure
*RST:
NONE
Example:
See
Specifying Measurement Control Settings
Firmware/Software:
V1.0.15.20
Manual operation:
See
CONFigure:FMSTereo:MEAS<i>:MOEXception
<MeasOnException>
Specifies whether measurement results that the R&S
CMW identifies as faulty or inac-
curate are rejected.
Parameters:
<MeasOnException> OFF | ON
ON:
Results are never rejected
OFF:
Faulty results are rejected
*RST:
OFF
Example:
See
Specifying Measurement Control Settings
Firmware/Software:
V1.0.15.20
Manual operation:
See
CONFigure:FMSTereo:MEAS<i>:MEValuation:SCOunt:RFCarrier
<StatisticCount>
Specifies the statistic count of the measurement. The statistic count is equal to the
number of measurement intervals per single shot. Use
CONFigure:...:MEAS<i>:...:REPetition SINGleshot | CONTinuous
to
select either single-shot or continuous measurements.
See also: "Statistical Settings" in the R&S
CMW user manual, chapter "Remote Con-
trol"
Measurement Control Commands