Command Reference
R&S
®
CMW-KM645
36
User Manual 1173.9740.02 ─ 04
●
40 (ARB File CRC Error)
:
The ARB file CRC check failed. The ARB file is corrupt and not reliable.
●
42 (ARB Header Tag Invalid)
:
The ARB file selected in the GPRF generator contains an invalid header tag.
●
43 (ARB Segment Overflow)
:
The number of segments in the multi-segment ARB file is higher than the allowed
maximum.
●
44 (ARB File not Found)
:
The selected ARB file could not be found.
●
45 (ARB Memory Overflow)
:
The ARB file length is greater than the available memory.
●
46 (ARB Sample Rate out of Range)
:
The clock rate of the ARB file is either too high or too low.
●
47 (ARB Cycles out of Range)
:
The repetition mode equals "Single Shot" and the playback length is greater than
40 s. Please reduce the playback length or set the repetition mode to "Continuous".
<Length> = (<Cycles> * <Samples> + <Additional Samples>) / <Clock Rate>
●
60 (Invalid RF-Connector Setting)
The individual segments of a list mode measurement with R&S CMWS use differ-
ent connector benches. This is not allowed. All segments must use the same
bench.
Check the "Info" dialog for the relevant segment numbers.
●
101 (Firmware Error)
:
Indicates a firmware or software error. If you encounter this error for the first time,
restart the instrument.
If the error occurs again, consider the following hints:
–
Firmware errors can often be repaired by restoring the factory default settings.
To restore these settings, restart your instrument and press the "Factory
Default" softkey during startup.
–
If a software package (update) has not been properly installed this is often indi-
cated in the "Setup" dialog, section "SW/HW-Equipment > Installed Software".
–
A software update correcting the error may be available. Updates are e.g. pro-
vided in the "CMW Customer Web" on GLORIS (registration required):
.
If you get firmware errors even with the properly installed latest software version,
please send a problem report including log files to Rohde & Schwarz.
●
102 (Unidentified Error)
:
Indicates an error not covered by other reliability values. For troubleshooting
please follow the steps described for "101 (Firmware Error)".
●
103 (Parameter Error)
:
Indicates that the measurement could not be performed due to internal conflicting
parameter settings.
A good approach to localize the conflicting settings is to start with a reset or preset
or even restore the factory default settings. Then reconfigure the measurement
step by step and check when the error occurs for the first time.
If you need assistance to localize the conflicting parameter settings, please contact
Rohde & Schwarz (see
http://www.service.rohde-schwarz.com