Command Reference
R&S
®
CMW-KM645
46
User Manual 1173.9740.02 ─ 04
Manual operation:
See
CONFigure:FMSTereo:MEAS<i>:MEValuation:AF:FILTer:WEIGhting
<Filter>
Selects the weighting filter for the AF analyzer input path.
Parameters:
<Filter>
OFF | AWEight
OFF:
Bypass weighting filter stage (filter off)
AWEight:
Select A-weighting filter
*RST:
OFF
Example:
See
Continuous and AF Measurements
Firmware/Software:
V1.0.15.20
Manual operation:
See
CONFigure:FMSTereo:MEAS<i>:MEValuation:AF:THDFrequency
<ThdFreqLeft>,
<ThdFreqRight>
Selects two reference frequencies for the AF Total Harmonic Distortion (THD), the
Total Harmonic Distortion and Noise (THD + N) / SINAD and the SNR measurements.
Parameters:
<ThdFreqLeft>
Range:
1 Hz to 10.5E+3 Hz
*RST:
1000 Hz
Default unit: Hz
<ThdFreqRight>
Range:
1 Hz to 10.5E+3 Hz
*RST:
1000 Hz
Default unit: Hz
Example:
See
Continuous and AF Measurements
Firmware/Software:
V1.0.15.20
Manual operation:
See
"AF > THD (N) Freq. Left / Right"
CONFigure:FMSTereo:MEAS<i>:MEValuation:AF:BANDpass
<CenterFrequency>,
<Bandwidth>
Defines the center frequency and bandwidth of the variable bandpass filter that the
R&S
CMW uses for AF measurements.
Parameters:
<CenterFrequency> Range:
20 Hz to 21E+3 Hz
*RST:
1000 Hz
Default unit: Hz
<Bandwidth>
Range:
10 Hz to 1000 Hz
*RST:
200 Hz
Default unit: Hz