![Rohde & Schwarz 1422.0703.02 User Manual Download Page 74](http://html.mh-extra.com/html/rohde-and-schwarz/1422-0703-02/1422-0703-02_user-manual_1477787074.webp)
Functional Check
R&S
®
RT-ZD01
31
User Manual 1422.0732.02 ─ 02
1. Set the attenuation ratio on the probe to 100:1.
2. Connect the R&S
RT-ZD01 probe to an R&S oscilloscope as described in
chapter 2.1, "Connecting the Probe to R&S Oscilloscopes"
3. Connect the red sprung hook to the square wave output of the oscillo-
scope.
4. Connect the black sprung hook to the probe ground connector of the oscil-
loscope.
5. Press the AUTOSET key on the oscilloscope.
A square wave with 1
V (pp) between 0
V and 1
V is displayed on the display.
www.allice.de
Allice Messtechnik GmbH