Product Description
R&S
®
RT-ZD01
11
User Manual 1422.0732.02 ─ 02
1 = Probe amplifier box
2 = Input leads
3 = Safety banana plug (4 mm)
4 = Sprung hooks
5 = Output lead
6 = BNC output connector
7 = Auxiliary ground connector
Probe amplifier box
Contains the high voltage divider, the active differential amplifier and other elec-
tronic components. All components are designed to ensure safe operation at haz-
ardous contact voltages within the specified working voltage and measurement
category. In particular, all air gaps and creeping distances comply with all current
safety standards to protect the user, the measurement object, and the probe
against any harm or damage.
The active differential amplifier takes the difference between the positive and neg-
ative signal input voltages. The probe transfers this difference signal to the oscil-
loscope. Common mode voltages are rejected.
The controls and indicators of the amplifier box are described in
Output lead
Connects the amplifier box to the oscilloscope or other measurement instruments
using a BNC connector. Its length of 90 cm allows for a comfortable working dis-
tance to the base unit. The output lead is specially qualified for high voltage
usage.
Input leads
Provide flexible contact to the DUT even in confined physical conditions. The
input leads are each 30 cm long, thus the maximum distance of the measurement
points is 60 cm. The 4 mm safety banana plugs can be used to contact the DUT
directly, or to connect suitable contact accessories like the sprung hooks deliv-
ered with the probe.