Characteristics of Differential Probes
R&S
®
RT-ZD01
25
User Manual 1422.0732.02 ─ 02
Fig. 3-2: Dynamic range and operating voltage window for both attenuation ratios 100:1
and 1000:1
The dependencies of dynamic range, operating voltage window and attenuation
ratio are shown in
. The dynamic range between the positive and nega-
tive signal pins depends on the selected attenuation, but the operating voltage
window between each of the signal pins and common ground is not affected by
the attenuation.
The
shows several examples for permissible and impermissible inputs.
Dynamic Range and Operating Voltage Window