Trying Out the Instrument
R&S
®
RTE
49
Getting Started 1326.1026.02 ─ 08
1. In the [HORIZONTAL] area, press the [ACQUISITION] key.
The "Horizontal" settings dialog box opens.
2. Set the "Average count" to
10
to perform 10 waveform acquisitions.
3. Close the "Horizontal" dialog box.
4. In the [TRIGGER] area, press the [RUN N× SINGLE] key.
10 waveform acquisitions are performed. The most recent acquisition is dis-
played in the diagram.
5. In the [ANALYSIS] area, press the [HISTORY] key.
The quick access "History" dialog box appears, and the history mode is on.
6. Tap "Play".
The 10 stored waveforms are displayed one after the other, but very fast.
7. In the "Current acq." field, enter
-4
to display the sixth waveform, counted from
acquisition start. The latest acquisition has the number 0, the oldest has -9.
8. Tap
to open the "History" setup dialog box.
9. Enter
-2
in the "Start acq" field. Tap "Newest" to enter
0
in the "Stop acq" field.
Thus the three latest acquisitions are displayed.
10. In the "Replay time per acq." field, enter
1 s
to display each waveform for one
second.
11. Enable the "Auto repeat" option to see the three waveforms repeatedly.
12. Tap "Play".
Displaying the Waveform History