R&S FSH
Putting into Operation
Configuring the R&S FSH
Quick Start Guide 1309.6269.12 - 09
32
1.6.1 Configuring the Hardware
The hardware settings contain settings that control internal and connected hardware.
Using auto accessory detection
If you are using any accessories while working with the R&S FSH, the R&S FSH is
able to identify the connected hardware.
► In the "Instrument Setup" dialog box, select the "Auto Accessory Detection" item.
► Press the ENTER key.
► Select the accessory detection state from the dropdown menu.
► Confirm the selection with the ENTER key.
When this feature is on, the name of the connected accessory is displayed in the
"Detected Accessory" field.
Configuring the BNC connectors
You can use the BNC connectors for various applications. For more information on the
supported applications, see "
► In the "Instrument Setup" dialog box, select the "BNC 1" item.
► Press the ENTER key.
A dropdown menu to select the BNC connector application opens.
The following applications are available:
-
Trigger input
-
Reference signal input
-
DC Bias Port 1 and Port 2
(Only with models .24 and .28.)
► Select the required application.
► Confirm the selection with the ENTER key.