Putting into Operation
Connectors on the R&S FSH
Quick Start Guide 1309.6269.12 - 09
22
1.4.2 Tracking Generator Output
The tracking generator output is located on the top of the R&S FSH.
The tracking generator output power of the R&S FSH is 0 dBm nominal.
With an integrated step attenuator, you can reduce the tracking generator output power
by a maximum of 40 dB in 1 dB steps.
Risk of instrument damage
To avoid damage to the tracking generator output, the reverse voltage up to the voltage
stated on the R&S FSH housing must not be exceeded.
1.4.3 Power Sensor Port
The connector for power sensors is located on the top of the R&S FSH.
The connector is configured especially for power sensors. It is used for both the power
supply and data transfer via the power sensor interface.
Additionally, you can connect accessories like the R&S TS-EMF isotropic antenna
(order no. 1158.9295.13) to the power sensor port.