Maintenance
R&S
®
TS-PCA3
54
User Manual 1152.3908.12 ─ 19
6
Maintenance
6.1
Important User Information
The Test System Versatile Platform R&S TS-PCA3 is maintenance free.
6.2
Cleaning
Elctrocution Hazard!
The R&S TS-PCA3 must powered down before starting with cleaning the device.
The following equipment and materials are recommended for cleaning the Test System
Versatile Platform R&S TS-PCA3:
●
Vacuum cleaner
●
Brush
●
Soft, lint-free cloths
Damaged Test System Versatile Platform
Never use aggressive products to clean the R&S TS-PCA3.
Electrical interfaces must not be cleaned with liquid products such as contact spray.
Depending on the environmental conditions, it may be necessary to remove the individ-
ual plug-in modules from the R&S TS-PCA3 and clean them with a vacuum cleaner.
Risk of electrostatic discharge (ESD)
Electrostatic discharge (ESD) can cause damage to the electronic components of the
instrument and the device under test (DUT). ESD is most likely to occur when you con-
nect or disconnect a DUT or test fixture to the instrument's test ports.
To prevent ESD, use a wrist strap and cord and connect yourself to the ground, or use
a conductive floor mat and heel strap combination.
Cleaning