48
CoaguChek Professional Monitoring Systems
Training Manual
Additional information on Error 406
Power the meter off and remove the test strip. Repeat the measurement using a new test strip
and blood taken from a new puncture site at the tip of another finger. Do not touch or remove
the test strip when a test in in progress.
Note on Error 406: The CoaguChek XS PT Test strip may be used for patients under a
combination therapy of oral anticoagulants plus heparin injections. For maximum heparin
concentrations which do not interfere with the test, refer to the package insert for the test strip.
Under no circumstances, however, should heparinized capillary tubes be used for sample
application. Be sure to apply the blood drop to the test strip within 15 seconds of lancing the
fingertip. In rare cases, patients with long clotting times (> 8 INR) may receive an “Error 406”
message on the meter display. If this error message appears again when the test is repeated,
the result must be checked using another method.
Always consult the CoaguChek XS Plus or Pro user manual, and contact Roche Technical
Support if you are unsure of how to correct an error, or if you are unsure if your CoaguChek is
working correctly.
Roche POC Technical Support tel: 0808 100 1920 (UK), 1 800 409 564 (Ireland)
Roche POC Technical Support email: [email protected]
Power the meter off and remove the test strip. Repeat the measurement using a new test strip and
blood taken from a new puncture site at the tip of another finger. Do not touch or remove the test
strip when a test in in progress.
Note on Error 406:
The CoaguChek XS PT Test strip may be used for patients under a
combination therapy of oral anticoagulants plus heparin injections. For maximum heparin
concentrations which do not interfere with the test, refer to the package insert for the test
strip. Under no circumstances, however, should heparinized capillary tubes be used for sample
application. Be sure to apply the blood drop to the test strip within
15 seconds
of lancing the
fingertip. In rare cases, patients with long clotting times (> 8 INR) may receive an “Error 406”
message on the meter display. If this error message appears again when the test is repeated, the
result must be checked using another method.
Always consult the CoaguChek XS Plus or Pro user manual, and contact Roche Technical Support if
you are unsure of how to correct an error, or if you are unsure if your CoaguChek is working correctly.
Roche POC Technical Support tel: 0808 100 1920 (UK), 1 800 409 564 (Ireland)
Roche POC Technical Support email: [email protected]
Summary of Contents for CoaguChek XS Plus
Page 1: ...CoaguChek Professional Monitoring Systems Training Manual...
Page 4: ......
Page 5: ...CONTACT AND ORDER DETAILS...
Page 6: ......
Page 8: ......
Page 9: ...SYSTEM INTRODUCTION...
Page 10: ......
Page 16: ......
Page 17: ...INTERFERENCES...
Page 18: ......
Page 22: ......
Page 23: ...TRAINING...
Page 24: ......
Page 28: ......
Page 30: ......
Page 34: ......
Page 36: ......
Page 37: ...CLEANING...
Page 38: ......
Page 44: ......
Page 45: ...TROUBLESHOOTING...
Page 46: ......
Page 49: ...APPENDIX 1...
Page 50: ......
Page 65: ...APPENDIX 2...
Page 66: ......
Page 67: ...Accuracy and precision in oral anticoagulation monitoring I know my value...
Page 77: ......